Evaluation of the appropriate time period between sampling and analyzing for automated urinalysis.

Dolscheid-Pommerich RC, Klarmann-Schulz U, Conrad R, Stoffel-Wagner B, Zur B

Publication: Biochem Med (Zagreb) (2016) • Vol 26 • Page 82-9PMID: 26981022View on PubMed

Purpose

This paper investigated the effects of storage duration on urine particle and dipstick analysis.

Conclusion

Preanalytical storage of urine affected the white blood cell, red blood cell, epithelial cell, and cast counts and significantly affected the misclassification rate for measurement of pH, leukocytes, nitrite, protein, ketones, urobilinogen, blood, and specific gravity. There was no affect of storage for up to 240 min prior to analysis on bacterial cells counts or the misclassification rate for glucose or bilirubin.

Studies (2)

Purpose

This study investigated the effects of storage duration on urine particle counts. Urine from 321 patients were selected for inclusion based on containing more than 20 erythrocytes or leukocytes per μL or more than 100 bacteria per μL. Analysis was conducted using a Sysmex UF-1000i flow cytometer.

Summary of Findings

Urine stored for 120 min or 240 min had significantly lower white blood cell counts (64 versus 72, P<0.001 and 57 versus 72, P<0.001, respectively), red blood cell counts (35 versus 39, P<0.001 and 27 versus 39, P<0.001, respectively), and casts (1 vs 1, P=0.009 and 1 versus 1, P<0.001, respectively) compared to urine analyzed within 90 min. Epithelial cell counts decreased when urine was stored for 240 min rather than less than 90 min (11 versus 11, P=0.041), but was not significantly different when stored for 120 min rather than less than 90 min. Conductivity was higher when specimens were stored for 120 or 240 min before analysis rather than less than 90 min (14.1 versus 14.0, P<0.001, and 14.8 versus 14.0, P<0.001). Bacterial counts were unaffected by preanalytical storage.

  • Biospecimens

    FluidUrine
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Other diagnoses
  • Platforms

    Cell count/volumeFlow cytometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageTime at room temperature
    120 min
    240 min
    <90 min