Skip to main content

Comparison of 21-gauge and 22-gauge aspiration needle during endobronchial ultrasound-guided transbronchial needle aspiration.

Nakajima T, Yasufuku K, Takahashi R, Shingyoji M, Hirata T, Itami M, Matsui Y, Itakura M, Iizasa T, Kimura H

Publication: Respirology (2011) • Vol 16 • Page 90-4PMID: 20920141View on PubMed

Purpose

This paper compared diagnosis rates and histological structure in matched 21-gauge (G) and 22-G endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA).

Conclusion

Although specificity and histological preservation were comparable, EBUS-TBNA biopsies obtained with a 21-G needle generally contained more tumor cells and blood contamination than those obtained with 22-G biopsies.  Further, 21-G biopsies had a slightly higher sensitivity than 22-G biopsies (100% versus 91.3%), but demonstrated a trend of decreased preservation of non-necrotizing granulomatous inflammation suggestive of sarcoidosis.

Studies (1)

Purpose

This study compared diagnosis rates and histological structure in matched 21G and 22-G EBUS-TBNA. EBUS-TBNA were obtained from 33 patients (45 lesions) by one surgeon, first using a 22-G needle and then a 21-G needle. During each pass, the needle was moved back and forth 10-15 times and passes were conducted until a core was available with a maximum of three passes. Cytological analysis was performed on the smears and histological analysis was conducted on the formalin-fixed core. When necessary for diagnosis, immunohistochemistry (IHC) was performed on formalin-fixed specimens but the antibodies used were not specified.

Summary of Findings

The number of adequate cells was greater in smears from 21-G than 22-G needles in 49.9% of biopsies (22 of 45) and fewer in only 24.4% of biopsies (11 of 45, P=0.0256); however, blood contamination was higher in 73.3% and lower in 6.7% of 22-G than 21-G biopsies (P=0.0001). The histological preservation was superior to the other needle gauge in a comparable number of 21-G and 22-G biopsies.  Further, the specificity of 21-G and 22-G needle biopsies was comparable (100%) but the sensitivity was slightly higher for 21-G than 22-G biopsies (100% versus 91.3%). Finally, there was a trend toward better preservation of non-necrotizing granulomatous inflammation suggestive of sarcoidosis in the 22-G than 21-G biopsies (62.5% better and 25% worse, P=0.0695).

  • Biospecimens

    TissueLung
    TissueLymph Node
    CellLung
    CellLymph Node
  • Preservation Types

    FormalinOther Preservative
  • Diagnoses

    NeoplasticBenign
    NeoplasticCarcinoma
    NeoplasticLymphoma
    Other diagnoses
  • Platforms

    MorphologyH-and-E microscopy
    MorphologyLight microscopy
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionNeedle gauge
    21-G
    22-G