This paper examined the effects of formalin-fixed paraffin-embedded (FFPE) section storage and DNA extraction method on DNA yield, purity, maximum amplicon length and the ratio of two PCR fragments.
Assessing quality and functionality of DNA isolated from FFPE tissues through external quality assessment in tissue banks.
Ahmad-Nejad P, Duda A, Sucker A, Werner M, Bronsert P, Stickeler E, Reifenberger G, Malzkorn B, Oberländer M, Habermann JK, Bruch HP, Linnebacher M, Schadendorf D, Neumaier M
Purpose
Conclusion
The maximum PCR fragment length and the ratio of a 379 to 152 bp fragment decreased with FFPE section storage. DNA yield and purity DNA yield, purity, maximum amplicon length and the ratio of 379/152 bp amplicons differed among case-matched FFPE sections extracted at different biobanks, likely due to the different extraction methods employed.
Studies (2)
Purpose
The purpose of this study was to compare DNA yield, purity, maximum amplicon length and amplicon ratio (379/152 bp) among samples extracted from the same FFPE tumor blocks by nine different laboratories. Each of the nine biobanks were sent three nonsequential 10 µM FFPE sections of lung, bile and stomach tumors and three pooled sections of small intestine tumors. To ensure uniformity between the specimens, serial sections were sent to different biobanks such that one biobank received sections 2, 12, and 22 and another sections 3, 13, and 23. Sections were shipped at ambient temperature, and DNA was extracted at the receiving biobank by their method of choice, and quantified by spectrophotometer before DNA samples were shipped back to the reference laboratory at room temperature, on ice packs or on dry ice. Upon arrival DNA concentrations and purity were determined by Nanodrop spectrophotometer and protein contamination was quantified using the BCA assay. Specimens were further analyzed by electrophoresis, multiplex PCR, and pyro sequencing of coagulation factor V (379 bp) and calcyphosine (152 bp). All specimens were compared to specimens extracted from the same block at the reference laboratory using the QIAamp DNA FFPE Tissue-Kit and to high molecular weight DNA extracted from buffy coat and degraded by incubation at 95˚C for 2.5-50 min.
Summary of Findings
DNA yield and purity differed among case-matched FFPE sections extracted at different biobanks, likely due to the different extraction methods employed. While higher or lower yields were consistently obtained from all specimens extracted by some biobanks in comparison to the reference lab, inconsistent results were obtained for others. Similarly, the maximum amplicon size obtained from FFPE specimens was influenced by the biobank that performed the extraction, with consistent amplification of a 411 bp fragment from samples extracted at some biobanks and unsuccessful amplification of a 105 bp fragment from specimens extracted by other biobanks. The ratio of 379/152 bp amplicons was also influenced by the biobank that performed the DNA extractions, with consistent and higher ratios obtained from specimens processed at some laboratories but not others. Although many extraction details were not described, the authors report that the highest quality DNA was obtained after samples were heated to 95˚C for 10 min during extraction, while the lowest quality was obtained after samples were heated to 95˚C for 40 min.
Biospecimens
TissueLungTissueStomachTissueSmall BowelFluidBilePreservation Types
FormalinDiagnoses
NeoplasticCarcinomaNeoplasticNot specifiedPlatforms
DNAElectrophoresisDNASpectrophotometryDNANext generation sequencingDNAPCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Analyte Extraction and Purification Analyte isolation method 8 unspecified methodsQIAamp DNA FFPE Tissue-KitPCR Specific Length of gene fragment 105 bp199 bp299 bp411 bpNext generation sequencing Specific Length of gene fragment 152 bp379 bp
Purpose
The purpose of this study was to determine the effects of FFPE section storage on maximum PCR amplicon length and 379/152 bp amplicon ratio. Thirty FFPE tissue sections stored between 4 and 9 years, (5 sections per each storage year investigated) were obtained from the melanoma tissue bank. However, details regarding storage condition and when they were sectioned were not provided.
Summary of Findings
The maximum PCR fragment length and 379/152 bp amplicon ratio declined with FFPE section storage, such that the 411 bp fragment could not be amplified from FFPE sections stored for 7 years prior to the publication of results. Further, the success rate for amplification of the 300 and 200 bp products also declined with FFPE section storage. The ratio of 379/152 bp amplicons was also influenced by storage of FFPE sections, ranging between 0.22-0.72 in sections stored for 5 years prior to the publication of results, 0.35-0.53 in sections stored for 7 years, 0.00-0.27 in sections stored for 9 years.
Biospecimens
TissueSkinPreservation Types
FormalinDiagnoses
NeoplasticMelanomaPlatforms
DNANext generation sequencingDNAPCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 4-9 yearsPCR Specific Length of gene fragment 105 bp199 bp299 bp411 bpNext generation sequencing Specific Length of gene fragment 152 bp379 bp