The purpose of this paper was to determine if the use of red Vacutainer tubes introduces bias when measuring protein in cerebrospinal fluid (CSF).
Use of red vacutainer for collection of CSF causes falsely high estimation of proteins by pyrogallol red method.
Amle DB, Koner BC
Purpose
Conclusion
Protein levels were significantly higher in specimens transferred to red Vacutainer tubes than in specimens that remained in the original collection tube. Importantly, the effect was systematic and occurred at all concentrations and in CSF both from neonates (<1 month) and older patients (>1 month).
Studies (1)
Purpose
The purpose of this study was to determine if the use of red Vacutainer tubes introduces bias when measuring protein in CSF from neonatal, pediatric, and adult patients. CSF from fifty patients was collected into plastic tubes and then 1 mL was aliquoted into a 4 mL red vacutainer tube. Protein concentrations were then determined by the pyrogallol red method.
Summary of Findings
Specimens that were transferred into red Vacutainer tubes prior to analysis had significantly higher protein levels than specimens that remained in the original plastic collection tubes (p<0.0001). Importantly, the effect was systematic and occurred at all concentrations and in CSF from both neonates (<1 month) and older patients (>1 month).
Biospecimens
FluidCerebrospinal FluidPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Patient age <1 month>1 monthStorage Type of storage container Plastic tubeRed Vacutainer