Use of red vacutainer for collection of CSF causes falsely high estimation of proteins by pyrogallol red method.

Amle DB, Koner BC

Publication: Clin Chim Acta (2015) • Vol 444 • Page 62-5PMID: 25668232View on PubMed

Purpose

The purpose of this paper was to determine if the use of red Vacutainer tubes introduces bias when measuring protein in cerebrospinal fluid (CSF).

Conclusion

Protein levels were significantly higher in specimens transferred to red Vacutainer tubes than in specimens that remained in the original collection tube. Importantly, the effect was systematic and occurred at all concentrations and in CSF both from neonates (<1 month) and older patients (>1 month).

Studies (1)

Purpose

The purpose of this study was to determine if the use of red Vacutainer tubes introduces bias when measuring protein in CSF from neonatal, pediatric, and adult patients. CSF from fifty patients was collected into plastic tubes and then 1 mL was aliquoted into a 4 mL red vacutainer tube. Protein concentrations were then determined by the pyrogallol red method.

Summary of Findings

Specimens that were transferred into red Vacutainer tubes prior to analysis had significantly higher protein levels than specimens that remained in the original plastic collection tubes (p<0.0001). Importantly, the effect was systematic and occurred at all concentrations and in CSF from both neonates (<1 month) and older patients (>1 month).

  • Biospecimens

    FluidCerebrospinal Fluid
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionPatient age
    <1 month
    >1 month
    StorageType of storage container
    Plastic tube
    Red Vacutainer