IDS iSYS automated intact procollagen-1-N-terminus pro-peptide assay: method evaluation and reference intervals in adults and children.

Morovat A, Catchpole A, Meurisse A, Carlisi A, Bekaert AC, Rousselle O, Paddon M, James T, Cavalier E

Publication: Clin Chem Lab Med (2013) • Vol 51 • Page 2009-18PMID: 24072575View on PubMed

Purpose

The purpose of this paper was to determine the stability and effects of patient age and gender, evaluation of serum versus plasma, specimen hemolysis and other potential interfering biomarkers, as well as the type of auto analyzer used on the quantification of intact procollagen-1-N-terminus propeptide (PINP) in specimens from Northern Europeans.

Conclusion

PINP measurements in serum or plasma were not affected by delaying centrifugation of blood by up to 3 days at room temperature, an additional 4 days of storage of plasma or serum at 4 degrees C, up to 5 freeze-thaw cycles of serum, or up to 133 days of storage of serum at -20 degrees C. However, storage of serum at -20 degrees C for 2.5 years resulted in significant increases in measured PINP. Overall, the Immunodiagnostic Systems (IDS) iSYS assay measured significantly higher PINP values than the Roche E170 assay. However, at both the low and high ends of the concentration ranges, the iSYS intact PINP assay measured lower values than the E170 total PINP assay. As hemoglobin concentration increased, PINP measurements by the IDS iSYS assay significantly declined. PINP concentrations were higher in plasma than in serum, but the difference was clinically insignificant. Gender and age specific differences as well as differences based on the city of collection were noted for PINP measurements in serum.

Studies (3)

Purpose

The purpose of this study was to determine the effects of storage temperature and duration and freeze-thaw cycling on the measurement of PINP by the IDS iSYS assay in specimens from healthy patients. 18 plasma specimens were used in plasma stability experiments, 5 serum and 4 plasma specimens were used in the centrifugation delay experiments, 23 serum specimens were used for the -20 degrees C storage experiments, and 10 serum specimens were used in the freeze-thaw cycling experiments. All specimens were stored at -20 degrees C prior to analysis.

Summary of Findings

When heparinized plasma was left at room temperature for 3 days and then refrigerated at 4 degrees C for 4 days, PINP decreased by an average of 5.5% compared to when the plasma was evaluated before the room temperature and refrigerated storage (p=0.0001), but this decrease was not greater than the between batch imprecision in the IDS iSYS PINP assay. PINP measurements in serum or plasma were not affected by delaying centrifugation of blood by up to 3 days at room temperature, an additional 4 days of storage of plasma or serum at 4 degrees C, up to 5 freeze-thaw cycles of serum, or up to 133 days of storage of serum at -20 degrees C (plasma not tested). However, storage of serum at -20 degrees C for 2.5 years resulted in significant increases in measured PINP (p<0.0001).

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    PeptideClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageFreeze/thaw cycling
    1 cycle
    3 cycles
    5 cycles
    StorageStorage duration
    110 d
    133 d
    2.5 y
    36 d
    4 d
    64 d
    87 d
    StorageStorage temperature
    -20 degrees C
    4 degrees C
    Room temperature
    StorageTime at room temperature
    0 d
    1 d
    3 d