Skip to main content

Peri-operative troponin monitoring using a prototype high-sensitivity cardiac troponin I (hs-cTnI) assay: comparisons with hs-cTnT and contemporary cTnI assays.

Lee GR, Jhanji S, Tarrant H, James S, Pearse RM, Fitzgibbon M

Publication: Ann Clin Biochem (2014) • Vol 51 • Page 258-68PMID: 24048718View on PubMed

Purpose

The purpose of this paper was to determine the effects of time of collection with respect to gastro-intestinal surgery, method of intravenous (IV) fluid treatment during surgery, patient age and gender, and type of auto analyzer on troponin concentrations in serum.

Conclusion

Patient age and gender only had significant influences on troponin concentrations measured by the high sensitivity cardiac troponin T (hs-cTnT) assay, but not the cardiac troponin I (cTnI) or hs-cTnI assays. Troponin concentrations increased in specimens collected post-operatively compared to pre-operative ones, regardless of IV fluid treatment, and significant increases were noted as early as immediately after surgery using the hs-cTnI assay. Further, troponin levels in specimens collected at 24 h post-op were higher from patients with than without post-operative complications, and this was particularly evident among the stroke volume (SV) guided fluid with dopexamine treatment group. The hs-cTnT assay measured a significantly higher number of troponin levels in the 99th percentile, regardless of time of collection or IV treatment group, but the authors note that the influences of patient age and gender on this assay are likely confounding variables.

Studies (1)

Purpose

The purpose of this study was to determine the effects of time of collection with respect to gastro-intestinal surgery, method of IV fluid treatment during surgery, patient age and gender, and type of auto analyzer assay on troponin concentrations in serum. Specimens were collected from 135 patients in serum separator tubes, centrifuged, and serum was stored at -70 degrees C until analysis. During surgery, IV fluid treatment was either guided by SV or central venous pressure (CVP).

Summary of Findings

Patient age and gender had significant influences on troponin concentrations measured by the hs-cTnT assay but not the cTnI or hs-cTnI assays, with patients who were 60 years or older showing more cTnT concentrations in the 99th percentile and a significantly higher median cTnT concentration in males than females (p<0.01). Troponin concentrations were higher in the 24 h post-operative specimens than in the pre-operative ones (p<0.05), regardless of IV fluid treatment, and significant increases were noted as early as immediately after surgery using the hs-cTnI assay. Further, troponin levels in specimens collected at 24 h post-op were higher from patients with than without post-operative complications, and this was particularly evident among the SV guided fluid with dopexamine treatment group. The hs-cTnT assay measured a significantly higher number of troponin levels in the 99th percentile, regardless of time of collection or IV treatment group, but the authors note that the influences of patient age and gender on this assay are likely confounding variables.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Other diagnoses
  • Platforms

    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    No post-operative complications
    Post-operative complications
    Underwent major gastrointestinal surgery
    PreaquisitionIntra-operative fluid administration
    CVP guided IV fluid
    SV guided IV fluid with dopexamine
    SV guided IV fluid without dopexamine
    PreaquisitionPatient age
    60 years or older
    < 60 years old
    PreaquisitionPatient gender
    Female
    Male
    Biospecimen AcquisitionTime of biospecimen collection
    24 h post-operatively
    8 h post-operatively
    Before gastrointestinal surgery
    Immediately after gastrointestinal surgery
    Clinical chemistry/auto analyzer SpecificTechnology platform
    ARCHITECT i1000SR hs-cTnI assay
    Access 2 cTnI assay
    Cobas hs-cTnT assay