Reference range of hemolysis index in serum and lithium-heparin plasma measured with two analytical platforms in a population of unselected outpatients.

Lippi G, Giavarina D, Gelati M, Salvagno GL

Publication: Clin Chim Acta (2014) • Vol 429 • Page 143-6PMID: 24362229View on PubMed

Purpose

The purpose of this paper was to determine the effects of analyzing serum versus lithium heparin plasma and type of auto analyzer on the measurement of hemolysis index (HI).

Conclusion

Significantly higher cell free hemoglobin levels were measured in serum than in plasma by the Roche Cobas c501 but not by the Siemens Dimension Vista 1500. Cell free hemoglobin values for both plasma and serum were slightly but significantly higher when measured with the Siemens Dimension Vista 1500 than with the Cobas c501.

Studies (1)

Purpose

The purpose of this study was to determine the effects of analyzing serum versus lithium heparin plasma and type of auto analyzer on the measurement of HI. Specimens were obtained from 135 outpatients. The order of draw between the serum and lithium heparin tubes was alternated. Serum and plasma were stored at -70 degrees C for one week before analysis.

Summary of Findings

When cell free hemoglobin was determined using the Cobas c501, significantly higher levels were measured in serum than in plasma. On the other hand, no significant differences in cell free hemoglobin were measured between serum and plasma using the Siemens Dimension Vista 1500. Cell free hemoglobin values for both plasma and serum were slightly but significantly higher when measured with the Siemens Dimension Vista 1500 than with the Cobas c501 (p<0.01 and p=0.02, respectively).

  • Biospecimens

    FluidSerum
    FluidPlasma
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
  • Platforms

    ProteinSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Spectrophotometry SpecificTechnology platform
    Roche Cobas c501
    Siemens Dimension Vista 1500