Skip to main content

Low volume tubes can be effective to reduce the rate of hemolyzed specimens from the emergency department.

Giavarina D

Publication: Clin Biochem (2014) • Vol 47 • Page 688-9PMID: 24594067View on PubMed

Purpose

The purpose of this paper was to determine the effects of using smaller tubes (decreased blood collection volume) on the incidence of hemolysis.

Conclusion

When specimens were collected in 2.5 mL tubes rather than 5.0 mL tubes, there was a reduction in the rate of specimen rejection for potassium measurement due to hemolysis from 8.7% (587/6,725) to 2.6% (505/19,318).

Studies (1)

Purpose

The purpose of this study was to determine the effects of using smaller tubes (decreased blood collection volume) on the incidence of hemolysis. The authors stated that specimens scoring 4 or more on the hemolysis index (0.50-2.00 g/L of cell-free hemoglobin) were rejected for potassium measurement and considered hemolyzed. 26,043 specimens were included in the study spanning 11 weeks of blood collection in 5.0 mL lithium heparin Vacuette tubes and 33 subsequent weeks of blood collection in 2.5 mL lithium heparin Vacuette tubes. The authors note that all specimens were sent through a 600 meter long pneumatic tube system.

Summary of Findings

When specimens were collected in 2.5 mL tubes rather than 5.0 mL tubes, there was a reduction in the rate of specimen rejection for potassium measurement due to hemolysis from 8.7% (587/6,725) to 2.6% (505/19,318).
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Cell count/volumeSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    2.5 mL Vacuette
    5.0 mL Vacuette
    Biospecimen Aliquots and ComponentsAliquot size/volume
    2.5 mL
    5.0 mL