The pneumatic tube system does not affect complete blood count results; a validation study at a tertiary care hospital.

Al-Riyami AZ, Al-Khabori M, Al-Hadhrami RM, Al-Azwani IS, Davis HM, Al-Farsi KS, Alkindi SS, Daar SF

Publication: Int J Lab Hematol (2014) • Vol 36 • Page 514-20PMID: 24373139View on PubMed

Purpose

The purpose of this paper was to determine if pneumatic tube system (PTS) transport effects complete blood count (CBC) results.

Conclusion

PTS transport resulted in a faster turnaround time for blood specimen analysis, yet CBC results in PTS-transported and courier-delivered specimens were generally very strongly correlated. However, in specimens from healthy individuals and those with thalassemia, basophil and monocyte counts were only modestly and strongly correlated, respectively, between PTS- and courier-transported specimens, and in patients with thalassemia, mean corpuscular hemoglobin (MCH) was strongly correlated between PTS- and courier-transported specimens.

Studies (1)

Purpose

The purpose of this study was to determine if PTS transport effects CBC results. Paired EDTA blood specimens from 59 healthy patients and 56 patients with thalassemia were sent 1565 feet to the lab by PTS at less than 6 m/s or they were sent in a cold box by courier and analyzed upon arrival. The mean turnaround time was 76 min for specimens transported by PTS and 128 min by courier.

Summary of Findings

Hemoglobin, hematocrit, mean corpuscular volume (MCV), red blood cell distribution width (RDW), red blood cell counts (RBC), white blood cell counts (WBC), reticulocytes, absolute neutrophils, lymphocytes, eosinophils and platelets (PLT) were very strongly correlated between PTS and courier transported specimens (r=0.9213-0.9958), regardless of diagnosis. PTS- and courier-transported specimens from healthy patients had only modestly correlated basophil counts (r=0.5758) and strongly correlated monocyte counts (r=0.8605). For specimens from patients with thalassemia, PTS- and courier-transported specimens similarly had modestly correlated basophil counts (r=0.4827) and strongly correlated monocyte counts (r=0.8867), but in addition, MCH was only strongly correlated (r=0.7642)

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Other diagnoses
    Normal
  • Platforms

    Cell count/volumeHematology/ auto analyzer
    ProteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Healthy
    Thalassemia
    StorageSpecimen transport duration/condition
    128 min
    76 min
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system