ANNALS EXPRESS: Pneumatic tube transport of blood stained cerebrospinal fluid specimens has no clinically significant effect on rates of haemolysis compared to manual transport.

Jones EM, Chesher D, Isbister G, Gillett MJ

Publication: Ann Clin Biochem (2016) • Vol 53 • Page 168-73PMID: 26071489View on PubMed

Purpose

The purpose of this paper was to determine the effect of pneumatic tube system (PTS) transport on hemolysis as measured by net oxyhemoglobin absorbance (NOA) in red blood cell (RBC)-spiked cerebrospinal fluid (CSF).

Conclusion

NOA increased with increased RBC content in CSF. Generally, transport method did not affect NOA, but when CSF was spiked with 3000 x 10^6 RBC/L, the NOA was significantly lower in hand-delivered than PTS-transported CSF.

Studies (1)

Purpose

The purpose of this study was to determine the effect of PTS transport on hemolysis as determined by NOA in CSF spiked with increasing concentrations of RBCs. CSF with <1 g/L oxyhemoglobin, bilirubin or protein was collected from an unspecified number of donors, pooled and stored at -20°C. The pooled CSF was aliquoted and spiked with capillary blood in saline to achieve the desired RBC content. Paired specimens were then transported by PTS or hand-delivered. Within 1 h of arrival, specimens were centrifuged, and supernatants were subsequently stored at 4°C for up to 6 h before analysis.

Summary of Findings

NOA increased with increased RBC content. When spiked with 3000 x 10^6 RBC/L the NOA was significantly lower in hand-delivered than PTS-transported CSF (p=0.036), but the difference was small (0.005 versus 0.008 absorbance units), and no transport-method dependent differences were noted at higher levels of RBC contamination.

  • Biospecimens

    FluidCerebrospinal Fluid
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
  • Platforms

    ProteinSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen components
    0 RBC/L
    3000 x 10^6 RBC/L
    6000 x 10^6 RBC/L
    7000 x 10^6 RBC/L
    8000 x 10^6 RBC/L
    9000 x 10^6 RBC/L
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system