Logistics and quality control for DNA sampling in large multicenter studies.

Nederhand RJ, Droog S, Kluft C, Simoons ML, de Maat MP

Publication: J Thromb Haemost (2003) • Vol 1 • Page 987-91PMID: 12871366View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage, mechanical disruption and processing delays due to specimen shipment on hemolysis, DNA yield and quality and amplification success from blood specimens.

Conclusion

DNA yield decreased with increasing storage of blood at room temperature but DNA yield was not affected by up to 28 days of storage at 4 or -20 degrees C. For specimens stored for >3 days, the average optical density (OD)260/280 ratios were 1.7 for specimens stored at room temperature, 1.8 for specimens stored at 4 degrees C, and 1.6 for specimens stored at -20 degrees C. The authors report PCR was successful from all specimens, regardless of storage duration or temperature. The authors report visible erythrocyte lysis with increasing room temperature storage or freezing. The authors state mechanical disruption resulted in hemolysis, and lower OD260/280 ratios like those observed in specimens stored at -20 degrees C. Processing delays due to mailing specimens led to increasing hemolysis.

Studies (2)

Purpose

The purpose of this study was to determine the effects of storage and mechanical disruption on the DNA yield, quality and amplification success from blood specimens. Blood was collected from 8 healthy individuals into K3EDTA tubes, and DNA was extracted by salting-out and stored at 4 degrees C. The PCR target was not specified. Mechanical disruption was conducted on a head-over vortex for 24 h or by mailing (unspecified distance, duration and temperature).

Summary of Findings

DNA yield decreased with increasing storage of blood at room temperature, with a 50% reduction observed by day 28 compared to day 0, but DNA yield was not affected by up to 28 days of storage at 4 or -20 degrees C. For specimens stored for >3 days, the average OD260/280 ratios were 1.7 for specimens stored at room temperature, 1.8 for specimens stored at 4 degrees C, and 1.6 for specimens stored at -20 degrees C with more variability as well, but significance was not stated. The authors report PCR was successful from all specimens stored at room temperature, 4 degrees C, and -20 degrees C. The authors report visible erythrocyte lysis increased with increasing duration of room temperature storage which resulted in a need for repeated lysis steps and increased pronase to lyse the leukocyte pellet, but less lysis was observed with storage at 4 degrees C. The authors report that exposure of the specimens to extreme conditions (head-over vortex for 24 h, or mail for 24 h) resulted in hemolysis and variable OD260/280 ratios like those observed after storage at -20 degrees C, but similar to specimens stored at -20 degrees C there was no change in DNA yields.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    DNAElectrophoresis
    DNAPCR
    DNASpectrophotometry
    Cell count/volumeMacroscopic observation
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsHemolysis
    Not induced
    Vortex-induced
    StorageBetween site transportation method
    Mailed
    Not transported
    StorageStorage duration
    0 days
    14 days
    21 days
    28 days
    3 days
    7 days
    StorageStorage temperature
    -20 degrees C
    4 degrees C
    Room temperature
    PCR SpecificLength of gene fragment
    1200 bp
    200 bp