Differential interferences of hemoglobin and hemolysis on insulin assay with the Abbott Architect-Ci8200 immunoassay.

Garinet S, Fellahi S, Marlin G, Capeau J, Lefèvre G, Bastard JP

Publication: Clin Biochem (2014) • Vol 47 • Page 445-7PMID: 24486652View on PubMed

Purpose

The purpose of this paper was to determine the effects of hemoglobin (Hb) levels, hemolysis and storage on the measurement of insulin in serum using Abbott Architect-Ci8200 immunoassay.

Conclusion

Regardless of initial insulin concentration, as hemolysate was added to specimens, the measured levels of insulin decreased, but the addition of purified Hb did not affect the measurement of insulin. Insulin levels decreased with room temperature storage of serum, and the addition of hemolysate to serum exacerbated this effect. Storage of serum at 4 degrees C led to smaller decreases in insulin levels than storage at room temperature.

Studies (1)

Purpose

The purpose of this study was to determine the effects of Hb levels, hemolysis and storage on the measurement of insulin in serum using the Abbott Architect-Ci8200 immunoassay. Serum was stored frozen for less than 1 month and pooled to obtain the relevant insulin concentrations. Hemolysate or purified Hb was added to obtain a range of experimental concentrations, and specimens were analyzed immediately.

Summary of Findings

Regardless of insulin concentration, as hemolysate was added to specimens, the measured levels of insulin decreased, but the addition of increasing levels of purified Hb did not affect the measurement of insulin. Insulin levels in non-hemolyzed serum decreased by 20% after 9 h of room temperature storage, but a 20% decrease was noted after just 1 h and 2 h in specimens with 148 mg/dL Hb and 80 mg/dL Hb, respectively. Incubation of non-hemolyzed specimens at 4 degrees C for 9 h resulted in a <10% decrease in insulin levels, but in hemolyzed specimens a 10% decline in measured insulin was observed after 3 h at 4 degrees C.
  • Biospecimens

    FluidSerum
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
  • Platforms

    ProteinChemiluminescence immunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    136 pmol/L insulin
    1378 pmol/L insulin
    265.5 pmol/L insulin
    43 pmol/L insulin
    610 pmol/L insulin
    Biospecimen Aliquots and ComponentsBiospecimen components
    0 mg/dL Hb
    150 mg/dL Hb
    200 mg/dL Hb
    420 mg/dL Hb
    650 mg/dL Hb
    80 mg/dL Hb
    Biospecimen Aliquots and ComponentsHemolysis
    Hemolysate added
    No hemolysate added
    StorageStorage duration
    0 h
    1 h
    2 h
    3 h
    4 h
    5 h
    6 h
    7 h
    8 h
    9 h
    StorageStorage temperature
    24 degrees C
    4 degrees C