The purpose of this paper was to determine the effects of hemoglobin (Hb) levels, hemolysis and storage on the measurement of insulin in serum using Abbott Architect-Ci8200 immunoassay.
Differential interferences of hemoglobin and hemolysis on insulin assay with the Abbott Architect-Ci8200 immunoassay.
Garinet S, Fellahi S, Marlin G, Capeau J, Lefèvre G, Bastard JP
Purpose
Conclusion
Regardless of initial insulin concentration, as hemolysate was added to specimens, the measured levels of insulin decreased, but the addition of purified Hb did not affect the measurement of insulin. Insulin levels decreased with room temperature storage of serum, and the addition of hemolysate to serum exacerbated this effect. Storage of serum at 4 degrees C led to smaller decreases in insulin levels than storage at room temperature.
Studies (1)
Purpose
Summary of Findings
Biospecimens
FluidSerumPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
ProteinChemiluminescence immunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level 136 pmol/L insulin1378 pmol/L insulin265.5 pmol/L insulin43 pmol/L insulin610 pmol/L insulinBiospecimen Aliquots and Components Biospecimen components 0 mg/dL Hb150 mg/dL Hb200 mg/dL Hb420 mg/dL Hb650 mg/dL Hb80 mg/dL HbBiospecimen Aliquots and Components Hemolysis Hemolysate addedNo hemolysate addedStorage Storage duration 0 h1 h2 h3 h4 h5 h6 h7 h8 h9 hStorage Storage temperature 24 degrees C4 degrees C