The purpose of this paper was to determine the effects of collection tube type on markers of hemolysis.
Comparison of BD Vacutainer® Rapid Serum Tube and plasma for haemolysis markers in the emergency department.
Ryan JB, Stuart LA, Southby SJ, Than MP, Mackay R, Florkowski CM, George PM
Purpose
Conclusion
With the exception of lactate dehydrogenase (LD), levels of hemolysis markers were lowest when blood was collected into gel-free heparin tubes and highest when collected into Vacutainer Rapid Serum Tubes (RST). Interestingly, a lower percentage of specimens collected in RST tubes had LD values above the reference range than specimens collected in Vacutainer plasma separator tubes with gel (PST) or non-gel heparin tubes, but no other significant differences were found in the percentage of specimens above the reference range.
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube type on markers of hemolysis. Blood specimens were collected from 347 patients in the emergency department via intravenous cannula. Blood was collected into the RST first followed by the plasma tubes. All specimens were centrifuged at 2800 g for 10 min.
Summary of Findings
Specimens collected in the gel-free heparin tube had significantly lower hemolysis indices, potassium, and phosphate levels than those collected in PST (p<0.0001, p<0.0001, and p=0.0314, respectively) or RST (p<0.0001, all) and also had lower magnesium and AST than those collected in RST (p<0.0001 and p=0.0054, respectively). Specimens collected in PST had lower hemolysis indices, potassium, phosphate, aspartate aminotrasferase (AST), magnesium and lactate dehydrogenase than those collected in RST (p=0.0003, p<0.0001, p<0.0001, p=0.0038, p<0.0001 and p<0.0001, respectively). Unlike the other markers of hemolysis, LD was lower in specimens collected in RST than in gel-free heparin tubes (p<0.0001) or PST (p<0.001), and was lower in specimens collected in gel-free heparin tubes than PST (p=0.0116). Further, a lower percentage of specimens collected in RST tubes had LD values above the reference range than specimens collected in PST (p<0.001) and non-gel heparin tubes (p=0.002). Although specimens collected in RST were more likely to have hemolysis indices, and phosphate, AST, and magnesium levels above the reference range than those collected in PST or gel-free heparin tubes, the differences were not significant.
Biospecimens
FluidBloodFluidPlasmaFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerCell count/volumeHematology/ auto analyzerProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Vacutainer PST IIVacutainer RSTVacutainer non-gel heparin TubeBiospecimen Aliquots and Components Blood and blood products PlasmaSerum