Technical reproducibility of single-nucleotide and size-based DNA biomarker assessment using DNA extracted from formalin-fixed, paraffin-embedded tissues.

Zhang S, Tan IB, Sapari NS, Grabsch HI, Okines A, Smyth EC, Aoyama T, Hewitt LC, Inam I, Bottomley D, Nankivell M, Stenning SP, Cunningham D, Wotherspoon A, Tsuburaya A, Yoshikawa T, Soong R, Tan P

Publication: J Mol Diagn (2015) • Vol 17 • Page 242-50PMID: 25746798View on PubMed

Purpose

The purpose of this paper was to determine the reproducibility of genotyping results obtained by the MALDI-TOF MS based MassARRAY iPLEX system using formalin-fixed paraffin-embedded (FFPE) specimens and copy number results by PCR in frozen and FFPE specimens.

Conclusion

FFPE specimen genotyping results for technical replicates were 100% concordant using the MassARRAY iPLEX system and 67.3-95.8% concordant, depending on target, by PCR. Copy number concordance between technical replicates for PCR targets glutathione S-transferase theta 1 (GSTT1) and the thymidylate synthetase (TYMS) 5'UTR was 100% for each target for the 8 frozen colorectal specimens.

Studies (1)

Purpose

The purpose of this study was to determine the reproducibility of genotyping results obtained by the MassARRAY iPLEX system using FFPE specimens and copy number results by PCR in frozen and FFPE specimens. Non-malignant blocks of FFPE tissue, from 3 clinical cohorts investigating effects of chemotherapy on gastric cancer, with average storage of 5 years, 12 years, and 16 years, were included in this study. DNA was extracted from selected regions of 5 deparaffinized 10µm sections using a modification of the QIAamp DNA micro and QIAamp FFPE tissue kits. DNA was extracted from 8 frozen colon specimens using an unspecified Qiagen kit.

Summary of Findings

Genotyping results obtained for 6 SNPs in 55 FFPE specimens were 100% concordant between 2 technical replicates using the MALDI-TOF MS-based MassARRAY iPLEX system. In contrast, genotyping results for copy number variants in GSTT1 and the TYMS 5'UTR obtained by PCR were only concordant between technical replicates for 33 of the 49 specimens (67.3%) and 46 of 48 specimens (95.8%), respectively. Concordance between technical replicates of the GSTT1 and the TYMS 5'UTR copy numbers, as determined by PCR, was 100% for each target for the 8 frozen colorectal specimens.

  • Biospecimens

    TissueStomach
    TissueEsophagus
    TissueLymph Node
  • Preservation Types

    FormalinFrozen
  • Diagnoses

    Normal
  • Platforms

    DNASNP assay
    DNAMALDI-TOF MS
    DNAPCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen PreservationType of fixation/preservation
    Formalin (buffered)
    Frozen
    MALDI-TOF MS SpecificTargeted nucleic acid
    DPYD rS1801159
    DPYD rs3918290
    ERCC1 rs11615
    ERCC1 rs3212986
    ERCC2 rs13181
    UMPS rs180101019
    MALDI-TOF MS SpecificTechnology platform
    PCR
    PCR SpecificTargeted nucleic acid
    GSTT1 deletion
    TYMS 5'-UTR 2R/3R repeat