Estimating the effects of hemolysis on potassium and LDH laboratory results.
Zou J, Nolan DK, LaFiore AR, Scott MG
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of mechanically induced hemolysis on measured plasma levels of LDH and potassium. A total of 26 specimens earmarked for routine testing were used in this study. Aliquots were aspirated through a 22 gauge needle 0, 3, 5, or 7 times to induced different levels of hemolysis.
Summary of Findings
On average, for every increase in the HI of 100, potassium levels increased by 0.5-0.8 mmol/l, but there were large inter-individual differences in this relationship (r=0.88). At the same time, far less inter-individual variation was seen in the relationship between the increases in HI and increases in LDH levels (r=0.98).
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinSpectrophotometryProteinClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Hemolysis Fine needle aspiration-inducedNot induced