Skip to main content

Multicentric validation of a rapid assay for heparin-induced thrombocytopenia with different specimen types.

Kolde HJ, Habrecht U, von Hesberg J, Panzer S, Seidel H, Bakchoul T, Sachs UJ

Publication: Blood Coagul Fibrinolysis (2014) • Vol 25 • Page 6-9PMID: 24365772View on PubMed

Purpose

The purpose of this paper was to determine the effects of using serum, citrated whole blood, EDTA anticoagulated whole blood and citrated plasma in testing for heparin-induced thrombocytopenia (HIT) using a rapid lateral flow immunoassay (LFI-HIT) and comparing results to those obtained with other assays.

Conclusion

The LFI-HIT assay results showed good agreement between serum, citrated whole blood, citrated plasma, and EDTA anticoagulated whole blood. The authors state that more false positive results were obtained using the ELISA and the particle gel immunoassay than the LFI-HIT. However, citrated whole blood yielded more false negative results in the LFI-HIT assay than the serum or plasma, and the authors do not recommend the use of whole blood for this assay.

Studies (1)

Purpose

The purpose of this study was to determine the effects of testing for HIT in serum, citrated whole blood, EDTA whole blood or citrated plasma using a LFI-HIT and comparing results to those obtained with other assays. Specimens were collected from patients with suspected HIT at 5 different medical centers using either Becton Dickinson tubes or Sarstedt tubes. One center included specimens from healthy volunteers, one center included EDTA anticoagulated whole blood specimens, and one center included serum specimens that had been frozen and thawed, and the authors cite unpublished data which determined that freezing had no effects on LFI-HIT results.

Summary of Findings

LFI-HIT results using citrated whole blood, citrated plasma, and EDTA anticoagulated whole blood were 93.6%, 96.7%, and 100% identical, respectively, to those obtained using serum. As expected, all specimen types from healthy volunteers were negative for HIT. The authors state that more false positive results were obtained using the ELISA and the particle gel immunoassay than the LFI-HIT. The LFI-HIT results agreed with the heparin-induced platelet aggregation assay (HIPA) results 92.7%, 92.9%, and 89.7% of the time for serum, plasma, and whole blood specimens, respectively. Specifically, citrated whole blood yielded more false negative results than the serum or plasma, and the authors do not recommend its use in the LFI-HIT assay.

  • Biospecimens

    FluidBlood
    FluidPlasma
    FluidSerum
  • Preservation Types

    None (Fresh)Frozen
  • Diagnoses

    Normal
    Other diagnoses
  • Platforms

    GlycoproteinImmunoassay
    GlycoproteinELISA
    Cell count/volumeFlow cytometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Healthy
    Suspected HIT
    PreaquisitionOther drugs
    Low molecular weight heparin
    Unfractionated heparin
    Biospecimen AcquisitionAnticoagulant
    Citrate
    EDTA
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Whole blood
    Biospecimen PreservationType of fixation/preservation
    Frozen
    None (fresh)
    Immunoassay SpecificTechnology platform
    ELISA
    HIPA
    LFI-HIT
    Particle gel immunoassay