Two site evaluation of the performance of a new generation point-of-care glucose meter for use in a neonatal intensive care unit.
Tendl KA, Christoph J, Bohn A, Herkner KR, Pollak A, Prusa AR
Purpose
The purpose of this paper was to determine the effects of medications and levels of other clinical chemistry analytes on the levels of glucose in blood from neonatal intensive care unit (NICU) patients using two different clinical chemistry analyzers.
Conclusion
Glucose levels measured by the StatStrip glucose meter or the ABL glucose meter were very strongly correlated throughout the range of levels, but there was a small negative bias for levels measured by the StatStrip device that increased with increasing glucose levels. Hematocrit, pH, partial pressure oxygen, sodium, or medication did not affect the accuracy of glucose measurements with StatStrip (ABL used as reference).
Studies (1)
Purpose
The purpose of this study was to determine the effects of medications and levels of other clinical chemistry analytes on the levels of glucose in blood from NICU patients using two different clinical chemistry analyzers. Blood was obtained by an unspecified method from 1-62 day old (23.14-41.29 weeks gestation) NICU patients. Blood was first assayed using an ABL analyzer and then with the StatStrip glucose meter.
Summary of Findings
Glucose levels measured by the StatStrip glucose meter or the ABL glucose meter were very strongly correlated (r2=0.975) throughout the range of levels with a mean bias of -4.52% for levels measured by the StatStrip device. The mean StatStrip bias was -4.03% when glucose levels were <75 mg/dL and -3.07% for specimens with <45 mg/dL glucose. Hematocrit, pH, partial pressure oxygen, sodium, or medication did not affect the accuracy of glucose measurements with StatStrip (ABL used as reference).
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Cardiovascular DiseaseOther diagnosesPlatforms
CarbohydrateClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerGasClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level 122-145 mmol/L sodium13-389 mg/dL glucose25-59 mm Hg PO225-77% hematocritpH of 7.15-7.39Preaquisition Diagnosis/ patient condition Child of diabetic motherCongenital heart diseaseCongenital infectionHyperbilirubinaemiaMeconium aspirationNecrotizing enterocolitisPrenatal asphyxiaRespiratory diseasePreaquisition Other drugs CaffeineParacetamolVitamin CClinical chemistry/auto analyzer Specific Technology platform ABL glucose meterStatStrip glucose meter