Influence of physical properties of cuvette surface on measurement of serum lipase.
Cattoir L, Pede V, Decavele AS, De Buyzere ML, Speeckaert MM, Seccombe D, Boterberg V, Dubruel P, de Bisschop F, Delanghe JR
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of type of assay and cuvette properties on the measurement of lipase in serum. Serum was analyzed immediately or stored frozen at -20 degrees C. Additional control serum specimens were purchased from Roche. The Beckman cuvettes were made of glass, the Roche cuvettes were made of a cyclo olefin polymer and the Siemens cuvettes were made of polymethyl methacrylate.
Summary of Findings
Serum lipase levels were significantly higher when measured using the Siemens dimension system than the Beckman or Roche systems (p<0.001). The authors note the cuvettes used for each assay were made of different materials, had different surface/volume ratios (Roche was highest and Siemens was lowest), different contact angles and different surface energies. Further, when the same reagents were used for all assays, the catalytic activity was higher using the Roche cuvettes than the Siemens or Beckman cuvettes (p<0.05). The lipase activity ratio was not correlated with bilirubin, alkaline phosphatase activity, HDL cholesterol, LDL cholesterol, total cholesterol, or triglycerides nor did it correlate with specimen surface tension.
Biospecimens
FluidSerumPreservation Types
None (Fresh)FrozenDiagnoses
Not specifiedPlatforms
ProteinColorimetric assaySteroidClinical chemistry/auto analyzerLipidClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Preservation Type of fixation/preservation FrozenNone (fresh)Storage Type of storage container Clyclo olefin polymer cuvetteGlass cuvettePolymethyl methacrylate cuvetteColorimetric assay Specific Technology platform Beckman Coulter SynchronRoche CobasRoche ModularSiemens Vista