Influence of physical properties of cuvette surface on measurement of serum lipase.

Cattoir L, Pede V, Decavele AS, De Buyzere ML, Speeckaert MM, Seccombe D, Boterberg V, Dubruel P, de Bisschop F, Delanghe JR

Publication: Clin Chem Lab Med (2013) • Vol 51 • Page 2109-14PMID: 23839813View on PubMed

Purpose

The purpose of this paper was to determine the effects of type of assay and cuvette properties on the measurement of lipase in serum.

Conclusion

Serum lipase levels and activity depended on the assay used. The lipase activity ratio was not correlated with bilirubin, alkaline phosphatase activity, high density lipoprotein (HDL) cholesterol, low density lipoprotein (LDL) cholesterol, total cholesterol, or triglycerides nor did it correlate with specimen surface tension.

Studies (1)

Purpose

The purpose of this study was to determine the effects of type of assay and cuvette properties on the measurement of lipase in serum. Serum was analyzed immediately or stored frozen at -20 degrees C. Additional control serum specimens were purchased from Roche. The Beckman cuvettes were made of glass, the Roche cuvettes were made of a cyclo olefin polymer and the Siemens cuvettes were made of polymethyl methacrylate.

Summary of Findings

Serum lipase levels were significantly higher when measured using the Siemens dimension system than the Beckman or Roche systems (p<0.001). The authors note the cuvettes used for each assay were made of different materials, had different surface/volume ratios (Roche was highest and Siemens was lowest), different contact angles and different surface energies. Further, when the same reagents were used for all assays, the catalytic activity was higher using the Roche cuvettes than the Siemens or Beckman cuvettes (p<0.05). The lipase activity ratio was not correlated with bilirubin, alkaline phosphatase activity, HDL cholesterol, LDL cholesterol, total cholesterol, or triglycerides nor did it correlate with specimen surface tension.

  • Biospecimens

    FluidSerum
  • Preservation Types

    None (Fresh)Frozen
  • Diagnoses

    Not specified
  • Platforms

    ProteinColorimetric assay
    SteroidClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen PreservationType of fixation/preservation
    Frozen
    None (fresh)
    StorageType of storage container
    Clyclo olefin polymer cuvette
    Glass cuvette
    Polymethyl methacrylate cuvette
    Colorimetric assay SpecificTechnology platform
    Beckman Coulter Synchron
    Roche Cobas
    Roche Modular
    Siemens Vista