Blood collection tube-related alterations in analyte concentrations in quality control material and serum specimens.

Bowen RA, Sattayapiwat A, Gounden V, Remaley AT

Publication: Clin Biochem (2014) • Vol 47(3) • Page 150-7PMID: 24240064View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection tube type on measurement of triiodothyronine (TT3), thyroxine (TT4), cortisol, and other clinical chemistry analyte levels in serum.

Conclusion

Statistically significant differences in TT4 were observed between the plain glass tubes (standard) and Vacuettes and serum separator tubes (SST), but the differences were not clinically significant. Levels of the other analytes were not significantly different between plain glass, Vacuette, plain plastic red-top tubes (PRT), rapid serum tubes (RST), and SST collection tubes.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection tube type on measurement of TT3, TT4, cortisol, and 14 other clinical chemistry analyte levels in serum. All serum was processed within 2 hours of collection. Serum in Vacuette, SST, and RST was left on the separator while serum in plain glass tubes and PRT was transferred to plastic test tubes after centrifugation. Specimens were either left at room temperature and tested within 4 h, or they were stored at 4 degrees C for up to 7 days prior to testing.

Summary of Findings

Statistically significant differences in TT4 were observed between the plain glass tubes (standard) and Vacuettes (p=0.0037) and SST (p=0.0001), but the differences were not clinically significant. Serum TT3 and cortisol measurements were not significantly different between plain glass, Vacuette, PRT, RST, and SST collection tubes. Nor were significant differences were observed between the 5 collection tubes types in the levels of 14 other clinical chemistry analytes in serum. The authors highlight that levels of TT3, TT4, and cortisol were significantly affected by collection tube type when QC material was poured into the tubes prior to analysis.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)Other Preservative
  • Diagnoses

    Normal
  • Platforms

    PeptideClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Glass tube
    PRT
    RST
    SST
    Vacuette