Influence of sample type and delayed separation from cells on the measurement of B-type natriuretic peptide with the Architect system.
Janssen MJ, Velmans MH, Willemse N
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of tube type, anticoagulant type, blood component measured, and room temperature delays before centrifugation on BNP levels measured by the Architect i2000. Specimens were obtained from patients admitted to a cardiology department and BNP levels ranging from 10-2600 ng/L. All specimens were frozen at -20 degrees C until assay. EDTA tubes with no gel separator and EDTA plasma preparation tubes (PPT) were used in the stability experiments.
Summary of Findings
Significantly lower levels of BNP were measured in serum from serum separator tubes (SST) than in EDTA plasma collected with no gel separator. Slightly but significantly higher levels of BNP were measured in lithium heparin plasma from plasma separator tubes (PST) than in EDTA plasma collected with no gel separator. There were no differences in BNP measurements in EDTA plasma from PPT and EDTA plasma collected with no gel separator. BNP levels were not significantly different when specimens collected in EDTA tubes without a gel separator or EDTA PPT were left at room temperature for up to 8 h before centrifugation. However, after 24 h at room temperature BNP levels showed significant decreases of -17.9% and -19.1% in regular EDTA and EDTA PPT, respectively.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
FrozenDiagnoses
Not specifiedCardiovascular DiseasePlatforms
PeptideClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant EDTALithium heparinNoneBiospecimen Acquisition Type of collection container/solution EDTA PPTEDTA tubes with no gel separatorLithium heparin PSTSSTBiospecimen Aliquots and Components Blood and blood products PlasmaSerumBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Time at room temperature 24 h30 min4 h8 h