Influence of sample type and delayed separation from cells on the measurement of B-type natriuretic peptide with the Architect system.

Janssen MJ, Velmans MH, Willemse N

Publication: Clin Chem Lab Med (2013) • Vol 51 • Page e235-7PMID: 23645124View on PubMed

Purpose

The purpose of this paper was to determine the effects of tube type, anticoagulant type, blood component measured, and room temperature delays before centrifugation on B-type natriuretic peptide (BNP) levels measured by the Architect i2000.

Conclusion

Heparin plasma and serum had significantly higher and lower levels of BNP, respectively, than EDTA plasma, but there were no significant differences between BNP levels measured in EDTA plasma from tubes with or without gel separators. Room temperature delays of up to 8 h prior to centrifugation did not affect BNP measurements in EDTA plasma from tubes with or without a gel separator, but decreases were noted after 24 h delays at room temperature.

Studies (1)

Purpose

The purpose of this study was to determine the effects of tube type, anticoagulant type, blood component measured, and room temperature delays before centrifugation on BNP levels measured by the Architect i2000. Specimens were obtained from patients admitted to a cardiology department and BNP levels ranging from 10-2600 ng/L. All specimens were frozen at -20 degrees C until assay. EDTA tubes with no gel separator and EDTA plasma preparation tubes (PPT) were used in the stability experiments.

Summary of Findings

Significantly lower levels of BNP were measured in serum from serum separator tubes (SST) than in EDTA plasma collected with no gel separator. Slightly but significantly higher levels of BNP were measured in lithium heparin plasma from plasma separator tubes (PST) than in EDTA plasma collected with no gel separator. There were no differences in BNP measurements in EDTA plasma from PPT and EDTA plasma collected with no gel separator. BNP levels were not significantly different when specimens collected in EDTA tubes without a gel separator or EDTA PPT were left at room temperature for up to 8 h before centrifugation. However, after 24 h at room temperature BNP levels showed significant decreases of -17.9% and -19.1% in regular EDTA and EDTA PPT, respectively.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
    Cardiovascular Disease
  • Platforms

    PeptideClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    EDTA
    Lithium heparin
    None
    Biospecimen AcquisitionType of collection container/solution
    EDTA PPT
    EDTA tubes with no gel separator
    Lithium heparin PST
    SST
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageTime at room temperature
    24 h
    30 min
    4 h
    8 h