The purpose of this paper was to determine the effects of blood collection tube volume on hemolysis.
Low volume tubes are not effective to reduce the rate of hemolyzed specimens from the emergency department.
Lippi G, Bonelli P, Graiani V, Caleffi C, Cervellin G
Purpose
Conclusion
A significantly lower rate of hemolysis was observed when serum was collected in 5.0 mL tubes than 3.5 mL tubes.
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood collection tube volume on hemolysis. Specimens were collected in a large urban emergency department into either 5.0 or 3.5 mL Vacutainer Serum Separator II Plus (SST II Plus) plastic serum tubes. Approximately 80% were collected via intravenous catheter while the remaining specimens were collected by straight needle venipuncture. The authors do not analyze these two collection methods separately.
Summary of Findings
5.2% of serum specimens collected in 3.5 mL SST II Plus were hemolyzed whereas only 3.5% were hemolyzed when specimens were collected in 5.0 mL SST II Plus plastic tubes (p<0.001).
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinSpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Aliquot size/volume 3.5 mL5.0 mL