Studies on the use of BD Vacutainer(R) SST II(TM) and RST(TM) in general practice: investigation of artefactual hyperkalaemia.

Huyghe T, Buntinx F, Bruyninckx R, Besard V, Vunckx J, Church S, Byron K, Rosa R, Blanckaert N

Publication: Ann Clin Biochem (2014) • Vol 51 • Page 30-7PMID: 23897107View on PubMed

Purpose

The purpose of this paper was to determine the effects of tube type, incomplete tube filling, tube inversion, specimen transport duration, and other collection related variables on hemolysis rates, potassium concentrations, and lactate dehydrogenase (LDH) activity in serum.

Conclusion

Measured serum LDH activity was significantly lower among specimens collected in Vacutainer Rapid Serum Tubes (RST) than those collected in Vacutainer Serum Separator II Advance Blood Collection Tubes (SST II). Slightly lower hemolysis indices were observed among specimens that were not inverted after collection than those inverted five times. Incomplete filling of the tube resulted in significantly higher measured potassium levels, LDH activity, and hyperkalaemia rates. Serum LDH activity was also significantly higher when specimens were collected at patients' homes than when collected at general practitioners' (GP) offices. Transport duration, diuretics or potassium-sparing diuretics, treatment with angiotensin-converting-enzyme (ACE)-inhibitors, and difficulty during blood draw did not affect potassium or LDH measurements.

Studies (1)

Purpose

The purpose of this study was to determine the effects of tube type (RST versus SST II), incomplete tube filling, tube inversion, specimen transport duration, and other collection related variables on hemolysis rates, potassium concentrations, and LDH activity in serum. While most specimens were drawn in one of two GP offices, 6% were obtained at the homes of the patients. Transport time of specimens to the laboratory for testing ranged from 10-265 min. Incomplete fill volumes were defined by the authors as less than 10% below the tube markings.

Summary of Findings

Tube type had no significant effects on serum potassium levels, rates of hyperkalaemia, or rates of elevated serum LDH. However, measured serum LDH activity was significantly lower among specimens collected in RST than those collected in SST II (p=0.02). Inverting the tubes 5 times versus 0 times did not significantly affect serum potassium, LDH activity, incidence of hyperkalaemia, or rates of elevated serum LDH. Slightly lower hemolysis indices were observed among specimens that were not inverted than those inverted 5 times (p=0.05). Incomplete tube filling resulted in significantly higher measured potassium levels (p<0.001), LDH activity (p<0.001), and hyperkalaemia rates that were 4.86 times higher than for tubes which were completely filled. Serum LDH activity was also significantly higher when specimens were collected at patients' homes than when collected at GP offices (p<0.001). Transport time, diuretics or potassium-sparing diuretics, treatment with ACE-inhibitors, and difficulty during blood draw did not affect potassium or LDH measurements.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    ProteinSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionOther drugs
    ACE-inhibitors
    Diuretics
    None
    Potassium-sparing diuretics
    Biospecimen AcquisitionLocale of biospecimen collection
    Collected in GP office
    Collected in patient's home
    Biospecimen AcquisitionMethod of fluid acquisition
    Difficult blood draw
    Not a difficult blood draw
    Biospecimen AcquisitionType of collection container/solution
    RST with thrombin clot activator
    SST II with silica clot activator
    Biospecimen Aliquots and ComponentsAliquot size/volume
    Complete fill volume
    Incomplete fill volume
    Biospecimen Aliquots and ComponentsBiospecimen mixing
    Tube inverted 5 times
    Tube not inverted
    StorageBetween site transportation method
    Unspecified transport method
    StorageSpecimen transport duration/condition
    10-265 min