Comparison between plasma and serum osteopontin levels: usefulness in diagnosis of epithelial malignant pleural mesothelioma.

Cristaudo A, Foddis R, Bonotti A, Simonini S, Vivaldi A, Guglielmi G, Ambrosino N, Canessa PA, Chella A, Lucchi M, Mussi A, Mutti L

Publication: Int J Biol Markers (2010) • Vol 25 • Page 164-70PMID: 20878622View on PubMed

Purpose

The purpose of this paper was to determine the effects of delayed centrifugation, storage of plasma or serum at -80 degrees C and freeze-thaw cycling of plasma and serum on levels of Osteopontin (OPN) in specimens from healthy individuals and those with benign respiratory disease or malignant pleural mesothelioma.

Conclusion

Plasma OPN levels were higher than serum OPN levels, but a weak correlation was observed. The authors report that delayed centrifugation for up to 1 h at room temperature or 15 min at room temperature and 360 min at 4 degrees C and up to 2 freeze thaw cycles did not impact plasma OPN levels. In contrast, serum OPN levels were affected by pre-centrifugation storage at room temperature but not at 4 degrees C and by up to 2 freeze-thaw cycles. The authors report that storage of plasma or serum at -80 degrees C for 2 weeks instead of 1 week resulted in significant declines in OPN levels.

Studies (1)

Purpose

The purpose of this study was to determine the effects of delayed centrifugation, storage of plasma or serum at -80 degrees C and freeze-thaw cycling of plasma and serum on levels of OPN in specimens from healthy individuals and those with benign respiratory disease or malignant pleural mesothelioma. To investigate the effects of delayed centrifugation, specimens were kept at room temperature for 15, 30 or 60 min, at 4 degrees C for 45 min, or at room temperature for 15 min followed by storage at 4 degrees C for 60, 90, 180 or 360 min before centrifugation. For all other studies, specimens were stored at room temperature for 15 min and 4 degrees C for 45 min before centrifugation. After centrifugation, plasma and serum were stored at -80 degrees C for 1 week, unless otherwise specified. Anticoagulant type was not specified.

Summary of Findings

Plasma OPN levels were higher than serum OPN levels, but a weak correlation was observed (R2=0.28, p<0.0001). The authors report that delayed centrifugation for up to 60 min at room temperature or 15 min at room temperature followed by up to 360 min at 4 degrees C did not impact plasma OPN levels, and serum OPN levels were not affected by pre-centrifugation storage at 4 degrees for up to 360 min, but serum OPN levels were reported to be affected by storage at room temperature for 60 min rather than 15 min. The authors do not report the values or direction of change. The authors also report that increased storage of plasma or serum at -80 degrees C resulted in significant declines in OPN levels. The authors state that serum OPN levels declined with each freeze-thaw cycle, but plasma OPN levels were stable for 2 freeze-thaw cycles. OPN levels in plasma and serum were affected by disease state.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
    NeoplasticBenign
    NeoplasticSarcoma
  • Platforms

    ProteinELISA
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Benign respiratory disease
    Healthy
    Malignant pleural mesothelioma
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage duration
    1 week
    105 min
    15 min
    195 min
    2 week
    3 weeks
    30 min
    375 min
    4 weeks
    45 min
    60 min
    StorageStorage temperature
    -80 degrees C
    4 degrees C
    Room temperature