Platelet function testing in hirudin and BAPA anticoagulated blood.
Mani H, Hellis M, Lindhoff-Last E
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of type of anticoagulant on platelet aggregation values and VASP phosphorylation after storage of blood specimens for up to 48 hours. After experimental room temperature storage, blood underwent centrifugation to obtain platelet rich plasma (PRP) immediately prior to platelet function testing.
Summary of Findings
Arachidonic acid (AA)-induced platelet aggregation values in PRP and AA-induced whole blood impedence aggregometry were not affected by up to 24 h and 12 h, respectively, of room temperature storage of hirudin or BAPA anticoagulated blood. However, with citrate as the anticoagulant, decreases in AA-induced platelet aggregation were seen for specimens stored at room temperature for 12 hours or more compared to 2 hours. Adenosine 5-diphosphate (ADP)-induced platelet aggregation values in PRP decreased after hirudin or BAPA anticoagulated blood were stored at room temperature for 24 hours compared to blood stored for 2 hours and in citrated blood stored for 12 hours at room temperature. ADP-induced whole blood impedance aggregometry showed significant declines in platelet responses when hirudin or BAPA anticoagulated blood from healthy volunteers was stored for 12 h compared to specimens stored for 2 hours (citrated blood was not tested), and results were lower in specimens from patients on clopidogrel, regardless of storage duration. The VASP assay was unaffected by storage of citrate, hirudin, or BAPA anticoagulated blood for up to 48 hours prior to analysis, regardless of diagnosis or drug therapy. However, a significantly higher platelet reactivity index was observed among hirudin and BAPA specimens that had been stored for 2 hours compared to citrated specimens that had been stored for 2 hours for both patients and healthy volunteers. The type of anticoagulant used resulted in significant differences in results of light induced transmission aggregometry after 2 hours of blood storage at room temperature prior to analysis that were agonist-, diagnosis-, and drug therapy-specific.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
None (Fresh)Diagnoses
NormalOther diagnosesPlatforms
Cell count/volumeHematology/ auto analyzerProteinFlow cytometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition HealthyPeripheral arterial occlusive diseasePreaquisition Other drugs AspirinClopidogrelNoneBiospecimen Acquisition Anticoagulant BAPAHirudinSodium citrateBiospecimen Aliquots and Components Blood and blood products Platelet-rich plasmaWhole bloodBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Time at room temperature 12 h2 h24 h48 hHematology/ auto analyzer Specific Reaction solution ADPArachidonic acid