Electrolyte balanced heparin may produce a bias in the measurement of ionized calcium concentration in specimens with abnormally low protein concentration.

Lyon ME, Guajardo M, Laha T, Malik S, Henderson PJ, Kenny MA

Publication: Clin Chim Acta (1995) • Vol 233 • Page 105-13PMID: 7758198View on PubMed

Purpose

The purpose of this paper was to determine the effects of using Radiometer Smooth-E arterial blood sampler syringes and incomplete filling on the measurements of electrolytes in whole blood and serum.

Conclusion

Ionized calcium, sodium, and potassium measurements in whole blood were not significantly affected by incomplete filling of Radiometer Smooth-E arterial blood sampler syringes or syringes with no electrolyte balanced heparin. Significant positive and negative biases were observed in measured ionized calcium concentrations when serum was spiked with <0.6 mM or >1.3 mM ionized calcium, respectively, and when the total protein concentration was 2.7 g/dL or less among specimens held in Radiometer Smooth-E arterial blood sampler syringes. The authors state that higher total protein concentrations eliminated ionized calcium concentration biases, and spiked serum protein and ionized calcium concentrations did not affect sodium or potassium measurements when specimens were held in Radiometer Smooth-E arterial blood sampler syringes.

Studies (2)

Purpose

The purpose of this study was to determine the effects of using Radiometer Smooth-E arterial blood sampler syringes and incomplete syringe filling on the measurements of electrolytes in whole blood. Radiometer Smooth-E arterial blood sampler syringes contained sodium and lithium heparin as well as calcium and potassium. Venous blood specimens were collected into these syringes and held in wet ice for 20 min prior to analysis or they were collected into plain plastic syringes (no heparin) and analyzed immediately.

Summary of Findings

Although smaller syringe fill volumes resulted in altered anticoagulant concentrations (40-240 units of heparin activity), ionized calcium, sodium, and potassium measurements in whole blood were not significantly affected by incomplete filling of Radiometer Smooth-E arterial blood sampler syringes. Further, they were not affected by incomplete filling of plain syringes with no heparin.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Lithium heparin
    None
    Sodium heparin
    Biospecimen AcquisitionType of collection container/solution
    Plastic collection syringe
    Radiometer Smooth-E arterial blood syringe
    Biospecimen Aliquots and ComponentsAliquot size/volume
    0.5 mL
    1.0 mL
    1.5 mL
    2.0 mL
    2.5 mL
    3.0 mL
    StorageStorage duration
    0 min
    20 min