Electrolyte balanced heparin may produce a bias in the measurement of ionized calcium concentration in specimens with abnormally low protein concentration.
Lyon ME, Guajardo M, Laha T, Malik S, Henderson PJ, Kenny MA
Purpose
Conclusion
Studies (2)
Purpose
Summary of Findings
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Lithium heparinNoneSodium heparinBiospecimen Acquisition Type of collection container/solution Plastic collection syringeRadiometer Smooth-E arterial blood syringeBiospecimen Aliquots and Components Aliquot size/volume 0.5 mL1.0 mL1.5 mL2.0 mL2.5 mL3.0 mLStorage Storage duration 0 min20 min
Purpose
The purpose of this study was to determine the effects of exogneous protein and calcium on the measurements of electrolytes in serum collected in using Radiometer Smooth-E arterial blood sampler syringes and spiked with various total protein and ionized calcium concentrations. Radiometer Smooth-E arterial blood sampler syringes contained sodium and lithium heparin as well as calcium and potassium. The pH of pooled patient serum that had been refrigerated at 4 degrees C for 7 days was adjusted to 7.4. Serum was deproteinized by ultrafiltration, spiked with various concentrations of bovine serum albumin and ionized calcium standards, aliquoted into either type of syringe, and held in ice water for 20 min prior to evaluation.
Summary of Findings
When serum specimens with 2.7 g/dL protein or less were held in Radiometer Smooth-E arterial blood sampler syringes, significant biases were observed in measured ionized calcium concentrations. Specifically, a positive bias occurred for serum spiked with <0.6 mM ionized calcium, and a negative bias occurred for serum spiked with >1 mM ionized calcium. The authors state that when total protein concentrations were >3.0 g/dL, ionized calcium concentration biases were not observed, regardless of the spiked concentration. The authors also state that spiked serum protein and ionized calcium concentrations did not affect sodium or potassium measurements when specimens were held in Radiometer Smooth-E arterial blood sampler syringes compared to those held in syringes with no electrolyte balanced heparin.
Biospecimens
FluidSerumPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Lithium heparinNoneSodium heparinBiospecimen Acquisition Type of collection container/solution Plastic collection syringeRadiometer Smooth-E arterial blood syringeBiospecimen Aliquots and Components Biospecimen components 0.25 to 2.0 mM ionized calcium0.8 g/dL protein2.0 g/dL protein2.7 g/dL protein>3.0 g/dL protein