Comparison of serum and heparinized plasma samples for measurement of chemistry analytes.

Miles RR, Roberts RF, Putnam AR, Roberts WL

Publication: Clin Chem (2004) • Vol 50 • Page 1704-6PMID: 15331516View on PubMed

Purpose

The purpose of this paper was to determine the effects of analyzing serum versus heparinized plasma on the measurements of 45 clinical chemistry analytes. For a subset of analytes, the results from two different analyzers were also compared.

Conclusion

Heparinized plasma had significantly lower levels of bile acids (-67%) and potassium (-6.0%) and higher levels of aldolase (+39%), ACE (+22%), and lactate dehydrogenase (+21%) than serum when measured on the Roche analyzer. Potassium and lactate dehydrogenase levels showed similar differences to those observed with the Roche analyzer between heparinized plasma and serum when measured by the Vitros 950 analyzer. However, the Vitros 950 analyzer also measured 20% higher bilirubin levels in plasma than serum. Importantly, the manufacturers of each analyzer state that both heparinized plasma and serum are acceptable for these assays where significant differences were found.

Studies (1)

Purpose

The purpose of this study was to determine the effects of analyzing serum versus heparinized plasma on the measurements of 45 clinical chemistry analytes. For a subset of analytes, the results from two different analyzers were also compared. For all twenty volunteers, serum tubes were filled prior to lithium-heparin tubes after a single venipuncture. Serum and plasma were stored at -70 degrees C for up to 8 months prior to analysis.

Summary of Findings

Heparinized plasma had significantly lower levels of bile acids (-67%) and potassium (-6.0%) and higher levels of aldolase (+39%), ACE (+22%), and lactate dehydrogenase (+21%) than serum when measured on the Roche analyzer. Potassium and lactate dehydrogenase levels showed similar differences to those observed with the Roche analyzer between heparinized plasma and serum when measured by the Vitros 950 analyzer. However, the Vitros 950 analyzer also measured 20% higher bilirubin levels in plasma than serum. Importantly, the manufacturers of each analyzer state that both heparinized plasma and serum are acceptable for these assays where significant differences were found.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    ProteinClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    LipoproteinClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot sequential collection
    1st collection
    2nd collection
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Roche Modular P analyzer
    Vitros 950 analyzer