Influence of two different buffered sodium citrate concentrations on coagulation testing.
Lippi G, Salvagno GL, Montagnana M, Guidi GC
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to compare PT, APTT, fibrinogen, and D-dimer concentrations between specimens from patients on oral anticoagulant therapy collected in plastic tubes containing 0.129 mol buffered sodium citrate (Grenier) and those collected in siliconized glass tubes containing 0.105 mol buffered sodium citrate (Becton-Dickinson).
Summary of Findings
Specimens drawn into plastic Grenier tubes containing 0.129 mol buffered sodium citrate had significantly longer PT and APTT and significantly lower fibrinogen concentrations than specimens drawn into siliconized glass Becton-Dickinson tubes containing 0.105 mol buffered sodium citrate (all p<0.01). The percentage biases for APTT and fibrinogen, but not PT, exceeded the guidelines for bias due to biologic variation. D-dimer concentration was not significantly different between specimens collected into the two tube types.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
MorphologyHematology/ auto analyzerGlycoproteinHematology/ auto analyzerPeptideELISAPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Multiple concentrations evaluatedSodium citrateBiospecimen Acquisition Type of collection container/solution Plastic tubeSiliconized glass tube