Skip to main content

Different manufacturers of syringes: a new source of variability in blood gas, acid-base balance and related laboratory test?

Lima-Oliveira G, Lippi G, Salvagno GL, Montagnana M, Picheth G, Guidi GC

Publication: Clin Biochem (2012) • Vol 45 • Page 683-7PMID: 22440459View on PubMed

Purpose

The purpose of this paper was to determine the effects of brand and type of heparin coated syringe used for collection on blood gas analysis.

Conclusion

Statistically significant differences were observed between specimens collected in four different types of heparin coated syringes for all 16 analytes investigated. Several of the differences observed were also clinically significant.

Studies (1)

Purpose

The purpose of this study was to determine the effects of brand and type of heparin coated syringe used for collection on blood gas analysis. The syringes compared were a calcium-balanced lithium heparin coated syringe (S-Monovette), an in-house prepared sodium heparin syringe (washed with 5000 I.U. sodium heparin), a spray-dried calcium-balanced lithium heparin syringe (BD Preset), and a lyophilized electrolyte-balanced lithium heparin syringe (Pico 50). The order of draw was randomized except the syringe containing in-house prepared sodium heparin was always last.

Summary of Findings

Statistically significant differences were observed between specimens collected in four different types of heparin coated syringes for all 16 analytes investigated. Several of the differences observed were also clinically significant including those between one or more syringe types for partial pressure carbon dioxide (pCO2), sodium, potassium, calcium, glucose, and lactate.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    GasClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Lithium heparin
    Multiple concentrations evaluated
    Multiple forms evaluated
    Multiple manufacturers evaluated
    Sodium heparin