The effect of recentrifugation of serum separator tubes on concentration of serum analytes.

Shafi H, Sadrzadeh H

Publication: Ann Clin Lab Sci (2012) • Vol 42 • Page 318-9PMID: 22964622View on PubMed

Purpose

The purpose of this paper was to determine the effects of recentrifugation of serum separator tubes (SST) following room temperature or refrigerated storage on levels of 10 clinical chemistry analytes.

Conclusion

When specimens were stored in the SST and recentrifuged prior to analysis, significantly higher levels of glucose, sodium, potassium creatinine, bicarbonate, phosphorus, and magnesium were measured than when analyzed immediately following the initial centrifugation.

Studies (1)

Purpose

The purpose of this study was to determine the effects of recentrifugation of SST following room temperature or refrigerated storage on levels of 10 clinical chemistry analytes. Blood was collected from 6 healthy individuals in SST. All tubes were centrifuged and assayed prior to room temperature or refrigerated storage and then again after storage. This study does not separately analyze the effects of post-centrifugation storage and a second centrifugation step.

Summary of Findings

After storage at room temperature for 2-24 h and recentrifugation, levels of glucose (p=0.003), sodium (p=0.005), creatinine (p<0.0001), bicarbonate (p=0.004), phosphorus (p=0.01), and magnesium (p=0.002) were higher than when assayed immediately after the initial centrifugation. Compared to baseline, specimens stored at 4°C for 2-24 h and recentrifuged had higher levels of potassium (p=0.004), creatinine (p<0.0001), bicarbonate (p=0.001), phosphorus (p=0.004), and magnesium (p=0.04) than when assayed immediately after the initial centrifugation. Blood urea nitrogen (BUN), chloride, and calcium were not affected by storage (either temperature) and recentrifugation.

  • Biospecimens

    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCentrifugation
    Different number of centrifugation steps compared
    StorageStorage duration
    0 h
    2 h
    24 h
    4 h
    8 h
    StorageStorage temperature
    4°C
    Room temperature