Skip to main content

Thrombin-accelerated quick clotting serum tubes: an evaluation with 22 common biochemical analytes.

Ng WY, Yeo CP

Publication: Adv Hematol (2013) • Vol 2013 • Page 769479PMID: 23653649View on PubMed

Purpose

The purpose of this paper was to determine whether blood collection into BD vacutainer rapid serum tubes (RST) rather than serum separator tubes (SST) effects the clotting time, results, or stability of 22 common clinical chemistry analytes during 4 degrees C storage.

Conclusion

Blood collected in RST showed an average clotting time of 2.07 min while specimens collected in SST required at least 15 min to clot. No statistically significant differences were found between specimens collected in RST and those collected in SST for any of the 22 analytes including a renal panel, a liver panel, a lipid panel, thyroid hormones, cardiac markers, and prostate-specific antigen. During refrigerated storage of serum collected in RST, no statistically significant changes were observed for the majority of analytes. However, during storage in RST, serum bicarbonate levels showed significant decreases, and sodium, potassium, chloride, and albumin showed statistically significant but clinically insignificant increases.

Studies (1)

Purpose

The purpose of this study was to determine whether blood collection into RST rather than SST effects the clotting time or the measurement or stability of 22 common clinical chemistry analytes during 4 degrees C storage. Specimens from healthy volunteers were centrifuged and loaded onto analyzers manually while those from patients were processed on an automated system which included centrifugation, aliquoting, and measurement in a linked analyzer. Specimens from patients were only tested for those analytes which were requested, but specimens from healthy volunteers were tested for all analytes in this study.

Summary of Findings

Blood collected in RST showed an average clotting time of 2.07 min while specimens collected in SST required at least 15 min to clot. No statistically significant differences were found between specimens collected in RST and those collected in SST for any of the 22 analytes including a renal panel, a liver panel, a lipid panel, thyroid hormones, cardiac markers, and prostate-specific antigen. During refrigerated storage of serum collected in RST, no statistically significant changes were observed for urea, creatinine, glucose, total protein, total bilirubin, alkaline phosphatase, alanine aminotransferase, aspartate aminotransferase, total cholesterol, high density lipoprotein (HDL)-cholesterol, triglycerides, thyroid stimulating hormone, free thyroxine, creatine kinase-MB, or N-terminal brain natriuretic peptide. Serum bicarbonate levels showed significant decreases during storage in RST, and sodium, potassium, chloride, and albumin showed statistically significant but clinically insignificant increases during storage in RST.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
    Normal
  • Platforms

    ProteinClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    MorphologyMacroscopic observation
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    RST
    SST
    StorageStorage duration
    0 d
    1 d
    2 d
    3 d
    4 d