Does laboratory automation for the preanalytical phase improve data quality?

Lima-Oliveira G, Lippi G, Salvagno GL, Danese E, Montagnana M, Brocco G, Voi M, Picheth G, Guidi GC

Publication: J Lab Autom (2013) • Vol 18 • Page 375-81PMID: 23686657View on PubMed

Purpose

The purpose of this paper was to determine the effects of manual versus automated preanalytical processing by the MODULAR PRE-ANALYTICALS EVO-MPA system as well as subsequent storage on routine clinical chemistry analytes in plasma.

Conclusion

Significant differences were noted for several clinical chemistry analytes between manual and automated processing regimes both immediately after processing and even more significant differences were noted between the two processing regimes when specimens were stored for 6 h after processing. Further, significant differences were observed in many clinical chemistry analytes when specimens that were manually processed were compared before and after 6 hours of storage at 4 degrees C. Storage of specimens processed in the MPA system in the MPA output buffer for 6 hours led to fewer significant changes in clinical chemistry analytes as compared to measurements before the additional storage.

Studies (1)

Purpose

The purpose of this study was to determine the effects of manual versus automated preanalytical processing by the MODULAR PRE-ANALYTICALS EVO-MPA system as well as subsequent storage on routine clinical chemistry analytes in plasma. After collection, all specimens were left at room temperature for 10 min. Paired specimens were then either centrifuged manually, uncapped manually, and loaded onto the analyzer, or they were loaded directly into the MODUALR PRE-ANALYTICALS EVO-MPA system where they were centrifuged and fed directly into the same analyzer as the manually processed specimens were analyzed with. After analysis, manually processed specimens were stored at 4 degrees C while those subjected to automated processing were stored in output buffer at room temperature in the MPA system. All specimens were reanalyzed after 6 h of storage.

Summary of Findings

Significant differences were noted for several clinical chemistry analytes between manual and automated processing regimes both immediately after processing and even more significant differences were noted between the two processing regimes when specimens were stored for 6 h after processing. Further, significant differences were observed in many clinical chemistry analytes when specimens that were manually processed were compared before and after 6 hours of storage at 4 degrees C. Storage of specimens processed in the MPA system in the MPA output buffer for 6 hours led to fewer significant changes in clinical chemistry analytes as compared to measurements before the additional storage.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    SteroidClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    Cell count/volumeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBlood processing method
    MODULAR PRE-ANALYTICALS EVO-MPA
    Manual
    Biospecimen Aliquots and ComponentsCentrifugation
    Automated
    Manual
    StorageShort-term storage solution
    MPA output buffer
    None
    StorageStorage duration
    0 h
    6 h
    StorageStorage temperature
    4 degrees C
    Room temperature