Phenindione interferes with measurement of creatinine.

Sankaralingam A, Karim Y, Swaminathan R

Publication: Clin Biochem (2013) • Vol 46(18) • Page 1912-3PMID: 24016729View on PubMed

Purpose

The purpose of this paper was to investigate whether the anticoagulant phenindione interferes with measurement of creatinine levels in serum using three different assay types.

Conclusion

When serum was spiked with clinically relevant concentrations of phenindione, there were no clinically relevant effects on measured levels of creatinine using LC-MS/MS or the Jaffe colorimetric assay. However, creatinine levels measured by the Roche enzyme assay decreased with increasing concentrations of phenindione resulting in clinically relevant discrepancies.

Studies (1)

Purpose

The purpose of this study was to investigate whether the anticoagulant phenindione interferes with measurement of creatinine levels in serum by LC-MS/MS, the Roche enzymatic assay or the colorimetric Jaffe method.

Summary of Findings

When serum was spiked with any concentration of phenindione, there were no clinically relevant effects on measured levels of creatinine using LC-MS/MS. However, creatinine levels measured by the Roche enzyme assay decreased with increasing concentrations of phenindione, and creatinine levels measured by the Jaffe colorimetric assay were elevated in specimens spiked with 10 g/L phenindione, but not less. The authors state the interference of phenindione with the Roche enzymatic assay produces clinically relevant discrepancies at clinically relevant concentrations of phenindione.

  • Biospecimens

    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Small moleculeColorimetric assay
    Small moleculeLC-MS or LC-MS/MS
    Small moleculeEnzyme assay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    281 umol/L creatinine
    30 umol/L creatinine
    80 umol/L creatinine
    Biospecimen Aliquots and ComponentsBiospecimen components
    1 g phenindione
    10 g phenindione
    100 mg phenindione
    2 g phenindione
    200 mg phenindione
    4 g phenindione
    400 mg phenindione
    5 g phenindione
    No phenindione
    Enzyme assay SpecificTechnology platform
    Jaffe method (colorimetric)
    LC-MS/MS
    Roche Enzyme assay