The purpose of this paper was to investigate whether the anticoagulant phenindione interferes with measurement of creatinine levels in serum using three different assay types.
Phenindione interferes with measurement of creatinine.
Sankaralingam A, Karim Y, Swaminathan R
Purpose
Conclusion
When serum was spiked with clinically relevant concentrations of phenindione, there were no clinically relevant effects on measured levels of creatinine using LC-MS/MS or the Jaffe colorimetric assay. However, creatinine levels measured by the Roche enzyme assay decreased with increasing concentrations of phenindione resulting in clinically relevant discrepancies.
Studies (1)
Purpose
The purpose of this study was to investigate whether the anticoagulant phenindione interferes with measurement of creatinine levels in serum by LC-MS/MS, the Roche enzymatic assay or the colorimetric Jaffe method.
Summary of Findings
When serum was spiked with any concentration of phenindione, there were no clinically relevant effects on measured levels of creatinine using LC-MS/MS. However, creatinine levels measured by the Roche enzyme assay decreased with increasing concentrations of phenindione, and creatinine levels measured by the Jaffe colorimetric assay were elevated in specimens spiked with 10 g/L phenindione, but not less. The authors state the interference of phenindione with the Roche enzymatic assay produces clinically relevant discrepancies at clinically relevant concentrations of phenindione.
Biospecimens
FluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Small moleculeColorimetric assaySmall moleculeLC-MS or LC-MS/MSSmall moleculeEnzyme assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level 281 umol/L creatinine30 umol/L creatinine80 umol/L creatinineBiospecimen Aliquots and Components Biospecimen components 1 g phenindione10 g phenindione100 mg phenindione2 g phenindione200 mg phenindione4 g phenindione400 mg phenindione5 g phenindioneNo phenindioneEnzyme assay Specific Technology platform Jaffe method (colorimetric)LC-MS/MSRoche Enzyme assay