Skip to main content

Quantification of Circulating miRNAs in Plasma: Effect of Preanalytical and Analytical Parameters on Their Isolation and Stability.

Sourvinou IS, Markou A, Lianidou ES

Publication: J Mol Diagn (2013) • Vol 15 • Page 827-34PMID: 23988620View on PubMed

Purpose

The purpose of this paper was to determine the effects of extraction method and storage of plasma on levels of spiked and endogenous microRNA (miRNA) by TaqMan real-time PCR.

Conclusion

The yields of cel-miR-39 and hsa-miR-21 were highest using the mirVana PARIS kit for extraction and lowest using TRIzol LS or the Norgen miRNA purification kit. Cycle threshold (CT) values for hsa-miR-21 and hsa-miR-16 increased with storage of plasma at 4, -20 or -70 degrees C, and significant increases, compared to 24 h of storage, were observed after 1 and 4 months at all three temperatures. However, the CT values increased by 3-4 cycles for hsa-miR-21 and hsa-miR-16 after storage of plasma at -20 or -70 degrees C, and the CT values increased by 14-15 cycles for hsa-miR-21 and hsa-miR-16 after storage of plasma at 4 degrees C for 4 months. Consequently, the authors conclude storage of plasma at -20 or -70 degrees C is better than storage of plasma at 4 degrees C. After extraction, cel-miR-39 was stable at -70 degrees C for at least 1 year.

Studies (2)

Purpose

The purpose of this study was to determine the effects of plasma volume, RNA precipitation temperature during TriZol LS extraction, and miRNA extraction kit type on the recovery of spiked cel-miR-39 and endogenous hsa-miR-21 in EDTA plasma from healthy individuals and patients with non-small cell lung carcinomas. EDTA blood was centrifuged and the was plasma spiked with cel-miR-39 before miRNA extraction.

Summary of Findings

Using the TriZol LS kit, the highest recovery of cel-miR-39 and hsa-miR-21 from plasma was achieved when the least volume of plasma was used (200 uL) and when RNA was precipitated at room temperature for 10 min rather than at -20 degrees C overnight. The yields of cel-miR-39 and hsa-miR-21 were highest using the mirVana PARIS kit for extraction and lowest using TRIzol LS (with optimized protocol) or the Norgen miRNA purification kit.
  • Biospecimens

    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
    NeoplasticCarcinoma
  • Platforms

    RNAReal-time qRT-PCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    200 uL
    250 uL
    300 uL
    500 uL
    Analyte Extraction and PurificationAnalyte isolation method
    Norgen miRNA purification kit
    TRIzol LS
    miRNeasy mini kit
    mirVana PARIS kit
    Analyte Extraction and PurificationAnalyte purification
    Precipitation at -20 degrees C
    Precipitation at room temperature
    Real-time qRT-PCR SpecificTargeted nucleic acid
    cel-miR-39
    hsa-miR-21