Random variation and systematic error caused by various preanalytical variables, estimated by linear mixed-effects models.

Sylte MS, Wentzel-Larsen T, Bolann BJ

Publication: Clin Chim Acta (2013) • Vol 415 • Page 196-201PMID: 23117035View on PubMed

Purpose

The purpose of this paper was to determine the effects of needle type and gauge, specimen mixing and pneumatic tube system (PTS) transport on levels of 22 clinical chemistry analytes in serum.

Conclusion

Specimens transported by PTS had significantly higher levels of lactate dehydrogenase (LD) and magnesium and higher standard deviation (SD) in the levels of creatinine kinase and glucose than specimens transported by hand-delivery. Serum from blood collected using 23-gauge butterfly needles had lower levels of calcium and creatinine kinase and higher SD in alkaline phosphatase (ALP) and LD than serum from blood collected through 21-gauge straight needles. Iron levels were higher when specimens were inverted only once rather than five times, but the SD and other analytes were not significantly affected.

Studies (2)

Purpose

The purpose of this study was to determine the effects of transporting blood specimens by PTS rather than hand-delivery on levels of 22 clinical chemistry analytes in serum. 4 blood specimens were collected from 60 non-fasting patients at the outpatient clinic. Blood was collected through 21-gauge straight needles into serum separator gel tubes which were inverted 5 times and stored for 10 min before transport. Prior to transport by PTS, specimens were wrapped in bubble wrap. Both transport by PTS and hand-delivery took 2.5 min.

Summary of Findings

Specimens transported by PTS had significantly higher LD (p<0.001) and magnesium levels (p=0.003) than specimens transported by hand-delivery. Further, the SDs in creatinine kinase and glucose levels were higher when specimens were transported by PTS than hand-delivered.

  • Biospecimens

    FluidBlood
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system