Evaluation of pneumatic-tube system for delivery of blood specimens.
Steige H, Jones JD
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of type of PTS, transport distance, tube fill volume, transport of unclotted blood and patient diagnosis on levels of clinical chemistry analytes in serum, plasma, and whole blood. Round-trip PTS transport involved transport over 2846 m, through 6 switches, and 134 bends, at a maximum speed of 9.75 m/sec. The wooden inserts tended to obstruct the tube. Urea and glucose were measured in oxalate-fluoride whole blood, fibrinogen was measured in oxalated plasma, and all other analytes were measured in serum. Significance of changes was not determined.
Summary of Findings
When whole blood was transported by PTS, serum levels of PO2, potassium, hemoglobin, and LDH levels were higher than in specimens that were not transported. When duplicate specimens were transported in wooden and vinyl inserts, potassium levels and hemoglobin levels were higher in specimens transported in wooden carriers than those transported in vinyl carriers, regardless of patient diagnosis. LDH and hemoglobin were also elevated in serum after PTS transport of unclotted blood, and the authors report similar findings after transport of heparinized blood. The increases in potassium, hemoglobin and LDH were greater when the tubes were only partially filled before PTS transport. When blood was transported one-way instead of round-trip, the increases in LDH and hemoglobin were smaller, and only half of the specimens had elevated potassium. Differences between transported and not-transported specimens in pH, sodium, phosphorus, chloride, calcium, partial pressure carbon dioxide, creatinine, uric acid, billirubin, total protein, fibrinogen and urea were within the analytical variability.
Biospecimens
FluidBloodFluidPlasmaFluidSerumPreservation Types
None (Fresh)Diagnoses
NormalNeoplasticLymphomaNeoplasticLeukemiaOther diagnosesPlatforms
CarbohydrateClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerGlycoproteinClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerGasClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Agnogenic myeloid metaplasiaChronic lymphocytic leukemiaHealthyMalignant lymphomaPolycythemia veraBiospecimen Acquisition Anticoagulant HeparinNoneOxalateBiospecimen Aliquots and Components Aliquot size/volume Tube filled 30%Tube filled 50%Tube filled 70%Biospecimen Aliquots and Components Blood and blood products PlasmaSerumWhole bloodStorage Specimen transport duration/condition One-way (1423 m)Round-trip (2846 m)Vinyl lined with sponge rubberWood insert with sponge rubber on bottomStorage Within hospital transportation method Not transportedPneumatic tube system