EFFECT OF PNEUMATIC TUBE TRANSPORT SYSTEM ON THE VALIDITY OF DETERMINATIONS IN BLOOD CHEMISTRY.
MCCLELLAN EK, NAKAMURA RM, HAAS W, MOYER DL, KUNITAKE GM
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of PTS transport on clinical chemistry analytes in blood and serum. Blood was collected into vacutainers which were sealed with rubber stoppers. During PTS transport, the tubes travelled at 25-30 ft/sec through several turns in carriers. The tubes were inserted into the carriers such that no tubes broke (stoppered end down). The significance of some of the reported changes was not evaluated.
Summary of Findings
The serum hemoglobin levels in the specimens transported via PTS were 145-4150% (mean 1528%) higher than when hand-delivered (p<0.0005). While none of the PTS-transported specimens had serum hemoglobin levels in the normal range, 14 of 18 hand-delivered specimens had levels in the normal range. Serum LDH levels were 21-163% (mean 70%) higher in specimens transported by PTS than those hand-delivered. Further, 16 of 18 specimens had higher serum potassium levels after PTS transport than after hand-delivery (2-32%). Serum levels of urea nitrogen, chloride, bicarbonate, sodium, glutamic oxalacetic transaminase, and glutamic pyruvic transaminase and blood levels of alkaline phosphatase, total protein, and albumin to globulin ratio were unaffected by PTS transport.
Biospecimens
FluidBloodFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinColorimetric assayProteinClinical chemistry/auto analyzerProteinSpectrophotometryElectrolyte/MetalFlame emission photometryElectrolyte/MetalClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Blood and blood products SerumWhole bloodStorage Within hospital transportation method Hand-deliveredPneumatic tube system