Validity of β-D-glucosidase activity measured in dried blood samples for detection of potential Gaucher disease patients.

Stroppiano M, Calevo MG, Corsolini F, Cassanello M, Cassinerio E, Lanza F, Stroppiana G, Cappellini MD, Filocamo M

Publication: Clin Biochem (2014) • Vol 47 • Page 1293-6PMID: 24945105View on PubMed

Purpose

The purpose of this paper was to determine if dried blood spots (DBS) can be used to identify Gaucher's disease based on β-glucosidase activity (GBA).

Conclusion

When DBS were used to diagnose Gaucher's disease, a cut-off of 2.75 pmol/punch/h rather than 4.4 pmol/punch/h resulted in higher specificity (94% versus 88.5%), but lower sensitivity (82.3% versus 88.2%) for identification of Gaucher's disease as confirmed in white blood cell extracts.

Studies (1)

Purpose

The purpose of this study was to determine if DBS can be used instead of white blood cell extracts to diagnose Gaucher's disease based on GBA.  The initial study involved 25 healthy individuals and 25 with Gaucher's disease. Validation was conducted with 443 DBS and 73 blood specimens. DBS were allowed to air-dry at room temperature overnight and were then stored at -20°C. DBS prepared elsewhere were shipped in sealed plastic bags.

Summary of Findings

The mean GBA was 7.9 pmol/punch/h (range 4.40-17.70) among 25 DBS from healthy individuals and 0.41 pmol/punch/h (range 0.03–1.10) among DBS from 25 individuals with Gaucher's disease. When a GBA cut-off value of 4.4 pmol/punch/h was applied to 443 DBS, 64 potential cases of Gaucher's disease were identified. Of these 64 cases, 49 were identified to be false positives using white blood cell extracts. The initial DBS screen did not identify Gaucher's disease in two patients who were found to have the disease based on white blood cell extracts. In one of these cases, a replicate DBS was below the cut-off, but for the other case the DBS GBA remained above the threshold. When a cut-off value of 2.75 pmol/punch/h was used instead of 4.4 pmol/punch/h, the specificity was higher (94% versus 88.5%), but the sensitivity was lower (82.3% versus 88.2%).

  • Biospecimens

    FluidBlood
    CellWhite Cells
  • Preservation Types

    Other PreservativeNone (Fresh)
  • Diagnoses

    Other diagnoses
    Normal
  • Platforms

    ProteinLowry protein assay
    ProteinFluorometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Gaucher's disease
    Healthy
    Biospecimen Aliquots and ComponentsBlood and blood products
    Leukocyte
    Whole blood
    Biospecimen PreservationType of fixation/preservation
    Air-dried
    None (fresh)