Skip to main content

Hepcidin-25: Measurement by LC-MS/MS in serum and urine, reference ranges and urinary fractional excretion.

Wolff F, Deleers M, Melot C, Gulbis B, Cotton F

Publication: Clin Chim Acta (2013) • Vol 423 • Page 99-104PMID: 23643854View on PubMed

Purpose

The purpose of this paper was to determine the effects of patient gender, patient age, selected biochemical analyte levels, absolute iron deficiency, anticoagulants, gel separators, and storage tube types on hepcidin-25 concentrations in serum and urine.

Conclusion

While collection of serum into serum tubes, with or without a gel separator, or lithium heparin tubes had no effects on hepcidin-25 levels, use of EDTA tubes led to the measurement of significantly higher hepcidin-25 levels. Storage tube material (plastic versus glass) had no effect on serum hepcidin-25 measurements, but storage of urine in plastic tubes led to a significant decrease in hepcidin-25 concentrations compared to storage of urine in glass tubes. While serum hepcidin-25 concentrations were significantly positively correlated with urine concentrations, creatinine, iron, transferrin saturation, hemoglobin, and ferritin, and levels were negatively correlated with transferrin. However, there were no correlations between serum hepcidin-25 levels and patient age or C-reactive protein levels. Serum hepcidin-25 levels were lower among patients with absolute iron deficiency than healthy individuals and among females than males. Recovery rates of hepcidin-25 from spiked serum were greater than from spiked urine, and urine levels of hepcidin-25 did not correlate well with ferritin levels.

Studies (1)

Purpose

The purpose of this study was to determine the effects of patient gender, patient age, selected biochemical analyte levels, absolute iron deficiency, anticoagulants, gel separators, and storage tube types on hepcidin-25 concentrations in serum and urine.

Summary of Findings

While collection of serum into serum tubes, with or without a gel separator, or lithium heparin tubes had no effects on hepcidin-25 levels, use of EDTA tubes led to the measurement of significantly higher hepcidin-25 levels (p<0.05). Storage tube material (plastic versus glass) had no effects on serum hepcidin-25 measurements, but storage of urine in plastic tubes led to a significant decrease in hepcidin-25 concentrations compared to storage of urine in glass tubes (p<0.05). While serum hepcidin-25 concentrations were significantly positively correlated with urine concentrations, creatinine, iron, transferrin saturation, hemoglobin, and ferritin, serum hepcidin-25 levels were negatively correlated with transferrin. However, there were no correlations between serum hepcidin-25 levels and patient age or C-reactive protein levels. Serum hepcidin-25 was lower among patients with absolute iron deficiency than healthy individuals (p<0.0001) and in females than males (p<0.05). Recovery rates of hepcidin-25 from spiked serum were greater than from spiked urine, and urine levels of hepcidin-25 did not correlate well with ferritin levels.

  • Biospecimens

    FluidSerum
    FluidUrine
    FluidBlood
    FluidPlasma
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
    Other diagnoses
  • Platforms

    ProteinLC-MS or LC-MS/MS
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    0.2-15 mg/L C-reactive protein
    12.1-15.9 g/dL hemoglobin
    20-343 ng/mL ferritin
    204-360 mg/dL transferrin
    47-180 ug/dL iron
    PreaquisitionDiagnosis/ patient condition
    Absolute iron deficiency
    No iron deficiency
    PreaquisitionPatient age
    25-65 years
    PreaquisitionPatient gender
    Female
    Male
    Biospecimen AcquisitionAnticoagulant
    EDTA
    Lithium heparin
    None
    Biospecimen AcquisitionBiospecimen location
    Plasma
    Serum
    Urine
    Biospecimen AcquisitionType of collection container/solution
    Serum tube with a gel separator
    Serum tube with no gel separator
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    StorageType of storage container
    Glass tube
    Plastic tube