Platelet concentrate transport in pneumatic tube systems--does it work?

Lancé MD, Marcus MA, van Oerle R, Theunissen HM, Henskens YM

Publication: Vox Sang (2012) • Vol 103 • Page 79-82PMID: 22220640View on PubMed

Purpose

The purpose of this paper was to determine the effects of specimen irradiation prior to pneumatic tube system (PTS) transport, PTS transport frequency and room temperature storage on markers of platelet (Plt) quality and function.

Conclusion

Storage of Plts at room temperature led to decreases in mean Plt counts, glucose content, and aggregation in response to collagen and adenosine diphosphate (ADP), and increases mean Plt volume (MPV), lactate content, percentage of Plts expressing CD62P, and microparticle counts, but room temperature storage of Plts did not affect pH. In contrast, irradiation of Plt specimens only led to decreased microparticle counts. While a single PTS transport on day 2 did not affect Plt quality or function, increased frequency of PTS transport led to decreased aggregation in response thrombin receptor activating-peptide (TRAP) as determined by multiple electrode aggregometry (MEA) and light transmission aggregation (LTA).

Studies (2)

Purpose

The purpose of this study was to determine the effects of irradiation prior to PTS transport and room temperature storage on markers of Plt quality and function, including aggregation induced by multiple methods. 12 Plt concentrates were transported by PTS on day 2 and 7 of storage. 6 of the specimens were irradiated before transport on day 2, and the other 6 remained non-irradiated. Plts were stored with constant agitation and assayed before and after irradiation and PTS transport on day 2 and again on day 7.

Summary of Findings

The mean Plt counts and glucose content decreased, and MPV, lactate content, percentage of Plts expressing CD62P and microparticle content increased between day 2 and 7 of room temperature storage of Plts (p<0.05, all), but pH was unaffected by storage. Further, Plt aggregation in response to collagen and ADP, as measured by both MEA and LTA, and aggregation in response TRAP by LTA decreased between day 2 and 7, but storage had no effects on Plt aggregation in response to arachidonic acid or TRAP by MEA or natriumarachidonic acid by LTA. Irradiation of Plts led to decreased microparticle counts (p<0.05), but it did not significantly affect any other Plt quality or function analytes. None of the analytes were significantly affected by PTS transport of Plts on either day.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Cell count/volumeHematology/ auto analyzer
    Cell count/volumeFlow cytometry
    Cell count/volumeELISpot
    ProteinClinical chemistry/auto analyzer
    ProteinELISA
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBlood processing method
    Irradiated
    Not irradiated
    StorageSpecimen transport duration/condition
    Irradiated
    Not irradiated
    StorageTime at room temperature
    2 days
    7 days
    StorageWithin hospital transportation method
    Not transported
    Pneumatic tube system
    Flow cytometry SpecificReaction solution
    ADP
    Arachidonic acid
    Collagen
    Natriumarachidonic acid
    TRAP
    Flow cytometry SpecificTechnology platform
    LTA
    MEA