Centrifugation protocols: tests to determine optimal lithium heparin and citrate plasma sample quality.

Dimeski G, Solano C, Petroff MK, Hynd M

Publication: Ann Clin Biochem (2011) • Vol 48 • Page 218-22PMID: 21398434View on PubMed

Purpose

The purpose of this paper was to determine the effects of centrifugation speed and duration and post-centrifugation storage on the measurement of potassium, lactate dehydrogenase (LD), glucose, and phosphorous in lithium heparin plasma and serum. In addition, the effects of centrifugation speed and duration were determined on prothrombin time (PT), activated partial thromboplastin time (APTT), derived fibrinogen and surface-activated clotting time (SACT) in citrated plasma.

Conclusion

Post-centrifugation storage of lithium heparin plasma for 6 days at 2-8°C led to increased potassium, LD, and phosphorous and decreased glucose, but storage of lithium heparin plasma for 6 h at 21°C only led to significantly increased LD. Storage of serum for 6 days at 2-8°C led to increases in phosphorous, but no other analytes were affected by post-centrifugation storage of serum, and increases in phosphorous were smaller than those observed after storage of lithium heparin plasma.

Centrifugation of citrate tubes for 3 min at 4000 g produced plasma with an elevated platelet count and significantly lower SACT than that observed after centrifugation for 10 min at 3000 g, but no other analytes were affected. The authors state that lithium heparin plasma quality increased with increasing centrifugation duration and speed, but the differences were less than the standard deviations.

Studies (2)

Purpose

The purpose of this study was to determine the effects of post-centrifugation storage on the measurement of potassium, LD, glucose, and phosphorous in lithium heparin plasma and serum. Blood was obtained from 14 healthy individuals.

Summary of Findings

Post-centrifugation storage of lithium heparin plasma for 6 days at 2-8°C led to increased potassium, LD, and phosphorous and decreased glucose, but storage of lithium heparin plasma for 6 h at 21°C only led to significantly increased LD. Storage of serum for 6 days at 2-8°C led to increases in phosphorous, but no other analytes were affected by post-centrifugation storage of serum, and increases in phosphorous were smaller than those observed after storage of lithium heparin plasma.

 

  • Biospecimens

    FluidPlasma
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    StorageStorage duration
    0 h
    6 days
    6 h
    StorageStorage temperature
    2-8°C
    21°C