Skip to main content

Pseudohyperkalemia due to pneumatic tube transport in a leukemic patient.

Kellerman PS, Thornbery JM

Publication: Am J Kidney Dis (2005) • Vol 46 • Page 746-8PMID: 16183430View on PubMed

Purpose

The purpose of this paper was to determine the effects of anticoagulant choice, collection container type and transport of blood specimens by pneumatic tube system (PTS) on potassium levels in plasma from patients with leukocytosis and patients with normal white blood cell (WBC) counts.

Conclusion

When blood from the patient with leukocytosis was transported by PTS in a lithium heparin vacutainer, 2-3 fold higher potassium levels were subsequently measured in the resulting plasma compared to when blood from the same patient was hand-delivered. However, PTS transport of whole blood had no effects on the levels of potassium in plasma from patients with normal WBC counts. Among specimens from the patient with leukocytosis, plasma potassium levels were unaffected by drawing blood into a syringe rather than a vacutainer or collection into lithium heparin versus heparin.

Studies (1)

Purpose

The purpose of this study was to determine the effects of anticoagulant choice, collection container type and transport of blood specimens by PTS on potassium levels in plasma from patients with leukocytosis and patients with normal WBC counts. Within a 25 min timeframe, blood was drawn from a single patient with leukocytosis into: Two lithium heparin vacutainers, one of which was transported by PTS, and another was hand-delivered, and three arterial blood gas syringes, two of which contained heparin, and one with lithium heparin, all of which were hand-delivered. Blood was collected from 5 patients with normal WBC counts and transported by PTS of hand delivered. No statistical analysis was presented.

Summary of Findings

When blood from the patient with leukocytosis was transported by PTS in a lithium heparin vacutainer, 2-3 fold higher plasma potassium levels were measured than when blood from the same patient was hand-delivered. However, PTS transport of whole blood had no effects on the levels of potassium in plasma from patients with normal WBC counts. Among specimens from the patient with leukocytosis, plasma potassium levels were unaffected by drawing blood into a syringe rather than a vacutainer or collection into lithium heparin versus heparin.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
    Other diagnoses
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Leukocytosis
    Patient with normal WBC count
    Biospecimen AcquisitionAnticoagulant
    Heparin
    Lithium heparin
    Biospecimen AcquisitionType of collection container/solution
    Arterial blood gas syringe
    Vacutainer
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system