The purpose of this paper was to determine the effects of aliquot volume and 1-2 years of frozen storage on plasma pyridoxal phosphate (PLP) levels.
A micromethod for determination of plasma pyridoxal phosphate and its use in assessment of storage stability of the vitamer.
Borschel MW, Kirksey A, Hamaker BR
Purpose
Conclusion
The mean PLP concentration among 12 fresh plasma specimens was 171.5 pmol/mL, but after 1-2 years at -30°C, the mean PLP concentration had declined to 153.1 pmol/mL. Levels of PLP before and after storage were modestly correlated, but when one specimen which had a large drop in PLP levels was excluded, the levels were very strongly correlated. Levels of PLP were comparable and very strongly correlated when 0.1 mL or 0.5-1 mL of plasma was used for measurement.
Studies (1)
Purpose
The purpose of this study was to determine the effects of aliquot volume and 1-2 years of frozen storage on plasma PLP levels. 12 plasma specimens with a range of PLP levels, but no specified diagnosis or anticoagulant, were analyzed before and after storage at -30°C.
Summary of Findings
The mean PLP concentration in the 12 plasma specimens prior to frozen storage was 171.5 pmol/mL, but after 1-2 years at -30°C, the mean PLP concentration had declined to 153.1 pmol/mL. Levels of PLP before and after storage were modestly correlated (r=0.67), but when one specimen which had a large drop in PLP levels was excluded, the levels were very strongly correlated (r=0.95). Levels of PLP were comparable and very strongly correlated (r=0.97) when 0.1 mL or 0.5-1 mL of plasma was used for analysis.
Biospecimens
FluidPlasmaPreservation Types
None (Fresh)FrozenDiagnoses
Not specifiedPlatforms
Small moleculeRadioassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Aliquot size/volume 0.1 mL0.5-1.0 mLStorage Storage duration 0 years1-2 years