Skip to main content

Laboratory-based ROTEM(®) analysis: implementing pneumatic tube transport and real-time graphic transmission.

Colucci G, Giabbani E, Barizzi G, Urwyler N, Alberio L

Publication: Int J Lab Hematol (2011) • Vol 33 • Page 441-6PMID: 21382181View on PubMed

Purpose

The purpose of this paper was to determine the effects of pneumatic tube system (PTS) transport on coagulation as determined by rotational thomboelastometry (ROTEM) analysis.

Conclusion

Extrinsic (EXTEM), intrinsic (INTEM), fibrin (FIBTEM) and aprotinin-inhibited (APTEM) thrombelastography parameters were not significantly affected by PTS transport of blood specimens rather than hand-delivery, and regardless of distance travelled, the values remained within the normal range for all specimens.

Studies (1)

Purpose

The purpose of this study was to determine the effects of PTS transport on coagulation as determined by ROTEM analysis. Two specimens were collected in each location which included the operating room, hematology clinic, and emergency room. Specimens were collected into tubes containing trisodium citrate and transferred at room temperature to the hematology laboratory. PTS transport from the hematology clinic to the laboratory was 1330 m, took 202.5 sec, and involved two transfer stations and was 440 m and took 300 sec by walking (p<0.001). PTS transport from the operating room to the laboratory was 420 m, involved one transfer station, and took 75 sec and was 160 m and took 107.5 sec by walking (p=0.034). PTS transport from the emergency room to the laboratory was 910 m, involved one transfer station, and took 155 sec and was 290 m and took 200 sec by walking (p<0.001). Once at the laboratory, specimens were analyzed immediately, in parallel on two machines. To eliminate bias based on machine, each machine alternated between specimens transported by PTS and those transported by hand.

Summary of Findings

None of the EXTEM, INTEM, FIBTEM and APTEM parameters were significantly affected by PTS transport rather than hand-delivery, and regardless of distance travelled, the values remained within the normal range for all specimens.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    MorphologyThromboelastography
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageSpecimen transport duration/condition
    1330 m in 202.5 sec
    160 m in 107.5 sec
    290 m in 200 sec
    420 m in 75 sec
    440 m in 300 sec
    910 m in 155 sec
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system