The purpose of this paper was to determine the effects of hemolysis, delayed freezing, and assay type on the quantification of insulin in EDTA plasma and serum.
Evaluation of immunochemiluminometric assays for the measurement of insulin and C-peptide using the ADVIA Centaur.
El Kenz H, Bergmann P
Purpose
Conclusion
Insulin levels declined with increasing hemolysate, reaching <50% of initial levels with 5% of red blood cells hemolyzed. The decline in insulin with increasing hemolysis was heightened when plasma or serum specimens were stored for 2 h before freezing. When assayed by chemiluminescent immunoassay instead of RIA, measured levels of insulin were 1.6 mU/L lower, and measured levels of c-peptide were higher with an increasing bias in c-peptide with increasing concentration.
Studies (1)
Purpose
The purpose of this study was to determine the effects of hemolysis, delayed freezing, and assay type on the quantification of insulin in EDTA-plasma and serum. The number of specimens used was not specified.
Summary of Findings
Insulin levels declined with increasing hemolysate, reaching <50% of initial levels with 5% of red blood cells hemolyzed. The decline in insulin with increasing hemolysis was heightened when plasma or serum specimens were stored for 2 h before freezing. When assayed by chemiluminescent immunoassay instead of RIA, measured levels of insulin were 1.6 mU/L lower, and measured levels of c-peptide were higher with an increasing bias in c-peptide with increasing concentration.
Biospecimens
FluidBloodFluidPlasmaFluidSerumPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
ProteinChemiluminescence immunoassayProteinRadioimmunoassayPeptideRadioimmunoassayPeptideChemiluminescence immunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Hemolysis AbsentPresentStorage Time at room temperature 0 h2 hChemiluminescence immunoassay Specific Technology platform RIA