The purpose of this paper was to determine the effects of hemolysis, storage, and method of analysis on the levels of insulin in EDTA plasma.
A mass spectrometric method for quantitation of intact insulin in blood samples.
Darby SM, Miller ML, Allen RO, LeBeau M
Purpose
Conclusion
Insulin levels declined by 57% and 24% over the 21 days of storage of plasma at 4°C and room temperature, respectively. Hemolysis did not affect the measurement of insulin by liquid chromatography (LC)-Mass spectrometry (MS), and levels of insulin levels measured by LC-MS were comparable to those obtained by radioimmunoassay (RIA).
Studies (1)
Purpose
The purpose of this study was to determine the effects of storing plasma at 4°C or room temperature and hemolysis on the measurement of insulin in EDTA plasma by LC-MS and to compare insulin levels measured by LC-MS and RIA. Pooled EDTA plasma stored under nitrogen was used to determine the stability of endogenous and spiked insulin. The number of specimens or pools used for each experiment was not specified.
Summary of Findings
Endogenous and spiked insulin levels declined steadily by 57% and 24% over the 21 days of storage of plasma at 4°C and room temperature, respectively. Specimens in which hemolysis was induced by fine needle aspiration had comparable insulin levels to non-hemolyzed specimens (1.3 versus 1.4 ng/mL). Insulin levels measured by LC-MS were comparable to those obtained by RIA. When not subjected to storage or hemolysis, 94% of the insulin spiked into whole blood was recovered in the plasma.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinLC-MS or LC-MS/MSProteinRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components 20 ng/mL insulin addedNo insulin addedBiospecimen Aliquots and Components Hemolysis Fine needle aspiration-inducedNot inducedStorage Storage duration 0 days14 days21 days3 days7 daysStorage Storage temperature 4°CRoom temperatureLC-MS or LC-MS/MS Specific Technology platform RIA