Evaluation of a dried blood spot assay to measure prenatal screening markers pregnancy-associated plasma protein a and free beta-subunit of human chorionic gonadotropin.

Cowans NJ, Suonpaa M, Kouru H, Wright D, Spencer K

Publication: Clin Chem (2013) • Vol 59 • Page 968-75PMID: 23426428View on PubMed

Purpose

The purpose of this paper was to determine the effects of using dried blood spots (DBS) instead of serum for analysis and temperature and humidity during DBS preparation on the measurement of human chorionic gonadotropin (hCG)-beta and pregnancy-associated plasma protein A (PAPP-A) in specimens from pregnant women.

Conclusion

The log base 10 serum and DBS concentrations of PAPP-A and hCG-beta were very strongly correlated, but the mean concentration PAPP-A was 10.2% lower while free hCG-beta was 82.9% higher in DBS than in serum. hCG-beta levels were within 10% of levels in DBS dried at 25 degrees C and 50% humidity when the drying temperature was between 25 and 32 degrees C and humidity was >20% or at all drying temperatures when the humidity was between 20% and 40%. PAPP-A levels in DBS were relatively unaffected by drying temperature, but they decreased by more than 10% when the humidity was greater than 65%.

Studies (1)

Purpose

The purpose of this study was to determine the effects of using DBS instead of serum and temperature and humidity during DBS preparation on the measurement of hCG-beta and PAPP-A in blood from pregnant women during the first trimester. Unless otherwise specified, DBS were prepared within 60 sec of venipuncture blood on Whatman protein saver collection cards, air dried at room temperature (25 degrees C) overnight with 50% humidity, and analyzed the following day. After spotting, the remaining blood was clotted, and serum was stored frozen at -20 degrees C until analysis.

Summary of Findings

The log base 10 serum and DBS concentrations of PAPP-A and hCG-beta were very strongly correlated (r=0.982 and r=0.959, respectively), but the mean concentration PAPP-A was 10.2% lower while free hCG-beta was 82.9% higher in DBS than in serum. Purification through an anti-hCG affinity column was responsible for loss of 1/3 of the hCG-beta but only a negligible amount of the PAPP-A. hCG-beta levels were within 10% of levels in DBS dried at 25 degrees C and 50% humidity when the drying temperature was between 25 and 32 degrees C and humidity was >20% or at all drying temperatures when the humidity was between 20% and 40%. PAPP-A levels in DBS were relatively unaffected by drying temperature, but they decreased by more than 10% when the humidity was greater than 65%.

  • Biospecimens

    FluidBlood
    FluidSerum
  • Preservation Types

    Other PreservativeNone (Fresh)
  • Diagnoses

    Pregnant
  • Platforms

    ProteinClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBlood and blood products
    Serum
    Whole blood
    Biospecimen PreservationType of fixation/preservation
    Air-dried
    None (fresh)
    StorageStorage conditions
    10-80% humidity
    50% humidity
    StorageStorage temperature
    20-40 degrees C
    Room temperature