Changes in concentration of DNA in serum and plasma during storage of blood samples.
Jung M, Klotzek S, Lewandowski M, Fleischhacker M, Jung K
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube type and storage of whole blood prior to centrifugation on levels of beta-globin DNA in plasma and serum from healthy blood donors. Plasma and serum were frozen at -80 degrees C until DNA extraction. DNA was extracted with NucleoSpin Blood DNA kit and stored at -20 degrees C until analysis.
Summary of Findings
Serum had 2.3-fold higher beta-globin concentrations than plasma (p<0.05), but the use of serum tubes containing kaolin-coated plastic granulate coagulation accelerator rather than plain serum tubes had no effect on beta-globin concentration. While pre-centrifugation storage of whole blood had no effect on beta-globin concentration in plasma, storage of whole blood for 2 h at room temperature prior to centrifugation led to significantly higher beta-globin concentrations in serum than when blood was centrifuged immediately (p<0.05). Further, a non-significant trend toward higher concentrations of beta-globin in serum was observed when whole blood was stored at 4 degrees C for 24 h prior to centrifugation compared to when specimens were centrifuged immediately.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
FrozenDiagnoses
NormalPlatforms
DNAReal-time qPCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Coagulation accelerator serum tubePlain serum monovettePotassium EDTA MonovetteBiospecimen Aliquots and Components Blood and blood products PlasmaSerumBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Storage duration 0 h2 h24 h4 h8 hStorage Storage temperature 4 degrees CRoom temperatureReal-time qPCR Specific Targeted nucleic acid Beta-globin