The purpose of this paper was to determine the effects of pneumatic tube system (PTS) transport of lithium heparin blood and analysis of serum rather than lithium heparin plasma on levels of hemolysis, creatinine kinase (CK), CK-muscle brain isoenzyme (CK-MB), and alanine aminotransferase (ALT).
Pneumatic tube-transported blood samples in lithium heparinate gel separator tubes may be more susceptible to haemolysis than blood samples in serum tubes.
Böckel-Frohnhöfer N, Hübner U, Hummel B, Geisel J
Purpose
Conclusion
When lithium heparin plasma specimens were transported by PTS rather than hand delivery, a higher percentage had a hemolysis index (HI) of >20. Lithium heparin plasma had significantly higher median HI, CK and CK-MB and lower median ALT than serum.
Studies (2)
Purpose
The purpose of this study was to determine the effects of PTS transport rather than hand delivery of whole blood on the HI of plasma. 1368 PTS-transported and 837 hand delivered lithium heparin plasma specimens from patients with unspecified diagnoses were included. All tubes were filled completely and sent at a speed of 4-6 m/s in a cartridge with a foam insert.
Summary of Findings
When lithium heparin plasma specimens were transported by PTS, 17.5% had a HI of >20, but only 2.6% of hand-delivered specimens had an HI of >20.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Cell count/volumeSpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Within hospital transportation method Hand-deliveredPneumatic tube system
Purpose
The purpose of this study was to determine the effects of analysis of serum rather than lithium heparin plasma on levels of hemolysis, CK, CK-MB, and ALT. 157 paired lithium heparin plasma and serum specimens with an HI of >20 after PTS were compared. All tubes were filled completely and sent at a speed of 4-6 m/s in a cartridge with a foam insert.
Summary of Findings
The median HI was higher in plasma specimens than serum specimens (85 versus 33, p<0.001). Further, plasma had significantly higher median CK and CK-MB and lower median ALT than serum (p<0.001, p<0.001, and p=0.017, respectively).
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Cell count/volumeSpectrophotometryProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Lithium heparin tubeSerum S-MonovettesBiospecimen Aliquots and Components Blood and blood products PlasmaSerum