Comparison of basophil activation test results in blood preserved in acid citrate dextrose and EDTA.
Sousa N, Martínez-Aranguren R, Fernández-Benitez M, Ribeiro F, Sanz ML
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of type of anticoagulant and storage at 4 degrees C on basophil activation test results in blood from 3 patients allergic to D pteronyssinus or P pretense. Buffy coat was used for analysis, and CD63 was used as the basophil activation marker.
Summary of Findings
Storage of blood specimens for 24 h at 4 degrees C prior to analysis did not affect basophil activation levels compared to when fresh blood was analyzed. A slight drop was noted in the percentage of activated basophils after blood was stored for 48 h at 4 degrees C, but the decreases were clinically irrelevant. There were no significant differences in basophil activation with regard to the use of K3EDTA or acid-citrate-dextrose. Further, basophil activation was not significantly different in response to anti-IgE receptor or higher or lower concentrations of allergen.
Biospecimens
FluidBloodFluidBuffy CoatPreservation Types
None (Fresh)Diagnoses
Other diagnosesPlatforms
Cell count/volumeFlow cytometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Allergic to D pteronyssinusAllergic to P pretenseBiospecimen Acquisition Anticoagulant Acid-citrate-dextrosePotassium EDTAStorage Storage duration 0 h24 h48 hFlow cytometry Specific Reaction solution Anti-IgE receptorHigher concentration of allergenLower concentration of allergenFlow cytometry Specific Targeted peptide/protein CD63