The purpose of this paper was to determine the effects of storage and patient gender on plasma viscosity.
Measurement of plasma viscosity by free oscillation rheometry: imprecision, sample stability and establishment of a new reference range.
Pedersen L, Nielsen EB, Christensen MK, Buchwald M, Nybo M
Purpose
Conclusion
Storage of plasma at 20 degrees C for 48 h, 4 degrees C for 20 days, or -20 degrees C for 3 months did not affect viscosity. There was no effect of patient gender on plasma viscosity.
Studies (1)
Purpose
The purpose of this study was to determine the effects of storage and patient gender on plasma viscosity. EDTA plasma from 20 healthy patients was stored at -20 degrees C for 3 months. After thawing specimens at 37 degrees C, plasma was then stored at 4 degrees C for 20 days, 20 degrees C for 2 days, or analyzed immediately. Reference ranges for plasma viscosity were based on plasma obtained from 585 Danish blood donors.
Summary of Findings
Storage at 20 degrees C for 48 h, 4 degrees C for 20 days, or -20 degrees C for 3 months did not affect plasma viscosity. There was no effect of patient gender on plasma viscosity.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
FrozenDiagnoses
NormalPlatforms
MorphologyRheometerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Patient gender FemaleMaleStorage Storage duration 2 days20 days3 monthsStorage Storage temperature -20 degrees C20 degrees C4 degrees C